Automated determination of urinary inorganic sulfates with the "AutoAnalyzer".
نویسنده
چکیده
A turbidimetric method for inorganic sulfates in urine, based on the precipitation of barium sulfate, is adapted to the AutoAnalyzer. Coating of the flowcell with BaSO4is avoided by an inter-sample air rinsing. Sampling rate is 40/h. Inorganic urinary sulfate determinations are of interest for the evaluation of endogenous acid production (1) and in experimental evaluation of the acid-excreting potential of the kidney (2-3). In earlier methods, urinary sulfate was determined gravimetrically as barium sulfate. Precipitation of sulfate, a method first used by Folin (4), is tedious and time-consuming. Subsequently, Rosenheim and Drummond (5) and Fiske (6) used a volumetric benzidine method, but interferences by chloride and phosphate seriously limited the usefulness of this technique (7). Finally, a nephelo-metric method was devised (8-10) and applied to automatic determinations of sulfate (11). This report deals with a modification of this latter technique and its adaptation to the N-methodology of Technicon. Rinsing the system with air prevents BaSO4 from coating the fi owcell. Manifold: (see Figure 1). Dialyzer: a single dialysis is performed at 37#{176}C, with a " C " membrane. Colorimeter: a 15-mm tubular flow cuvet and a 520-nm filter are used. Sampling: the sampling rate is 40/h. Reagents Diluent: Distilled water plus " Brij 35 " (1 ml/-liter water). HC1, iN: To 80 ml of concentrated HC1 in a volumetric flask, add water to 1 liter. BaCl2-gelatin reagent: Add 2 g of gelatin and 0.1 g of thymol to about 500 ml of distilled water and heat while stirring until dissolved. Cool to room temperature, filter through no. 42 Whatman filter paper, and add 20 g of barium chloride slowly with constant stirring. Dilute the mixture to 1 liter with distilled water. Procedure The main difficulty was to prevent the coating of the cuvet walls with BaSO4. Rinsing with DTA or H3P04 aspirated between each sample is inefficient. Instead, air rinsing was found to be quite adequate. As shown in Figure 1, a Technicon three-way valve was inserted in the manifold between the last mixing coil and the colorimeter. At regular intervals, the valve deviates the barium sulfate liquid suspensions from the colorimeter to the waste line (valve-way 2 in Figure 1), while air is aspirated for 22 s through the cuvet by means of an AO glass fitting adapted on the upper part of the T-debubbler. Subsequently, sample and reagent flow resume in the tubular cell while air flow is …
منابع مشابه
An automated distillation method for the determination of diacetyl in beer: a comparison of analysis by AutoAnalyzer and gas chromatography
was the case with the first design. Another advantage of the procedure described is that it enables the determination of total carbon and inorganic carbon in the range 10 to 50 mg/dm at a rate of 20 samples/h at a standard deviation of less than per cent. The described automated system is used very effectively in the authors laboratory for routine sample analysis. ACKNOWLEDGEMENTS This work wil...
متن کاملAutomated determination of urinary creatinine without sample dilution: theory and practice.
The rate of the Jaffé reaction depends on the concentration of sodium hydroxide; the pseudo-first-order rate constant of the reaction, at 37 degrees C in 10 mmol/L picrate solution, is 0.004 mmol/L. We formulated an automated method to determine urinary creatinine directly without manual sample dilution. The conditions are as follows: 10 mmol/L picrate and 60 mmol/L sodium hydroxide (final conc...
متن کاملTable 2 . Precision of Glucose Determination by the Cupric - Neocu proine and Glucose Oxidase
Methodological modifications have been made in our “SMA 12/60 AutoAnalyzer” system, to improve the reproducibility and accuracy of six commonly performed procedures. These include glucose with the glucose oxidase method, albumin with the bromcresol green method, and alkaline phosphatase with the sodium thymolphthalein monophosphate method. We have also modified the composition of the reagents u...
متن کاملDetermination of serum or plasma glucose on the "AutoAnalyzer II" by use of the hexokinase reaction.
Manual procedures based on these reactions and involving the direct addition of serum to a single reagent are commonly used as emergency procedures in routine clinical chemistry laboratories of hospitals. We have adapted one of these methods, “Glucose 500-Pack” (Calbiochem, P.O. Box 12087, San Diego, Calif. 92772) to an automated method on the “AutoAnalyzer II” (Technicon Instruments Corp., Tar...
متن کاملA simplified method for automated phosphorus analysis of serum and urine.
A simplified procedure utilizing ferrous ammonium sulfate-thiourea as the reductant for the determination of inorganic phosphate in serum and urine by the AutoAnalyzer is described. Elimination of the heating bath usually required for many phosphorus methods for color development yields an improved S.D. of ± 0.25 and facilitates troubleshooting during any apparatus malfunction. Comparison of re...