Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS
ثبت نشده
چکیده
This application note describes the quantification of cadmium, chromium, lead and mercury in plastic materials by external calibration using the Agilent 7500ce ICP-MS system equipped with an Octopole Reaction System (ORS). Isotope dilution analysis with ICP-MS was used as a confirmatory technique. Good agreement of the results obtained by the two calibration approaches was achieved. The sample digestion methodology, tuning considerations, results and uncertainty estimations are discussed. The method validated with IDMS results has potential as a fast analytical tool for compliance testing laboratories. Author Sarah Hill and Rebeca Santamaria-Fernandez LGC Ltd. Queens Road Teddington, Middlesex, TW11 0LY UK Application note
منابع مشابه
Multi-Element Determination of Heavy Metals in Dietary Supplements Using Collision/Reaction Cell ICP-MS Application
Determination of 11 metals (including arsenic, chromium, mercury, iron, copper, nickel, zinc, selenium, lead, cadmium, and thallium) in UK-consumed dietary supplements was carried out using ICP-MS. The instrument used was the Agilent 7500ce, which is equipped with a collision/ reaction cell (Octopole Reaction System), and was operated in no-gas, helium, and hydrogen modes, all acquired within a...
متن کاملRoHS/ELV Directives - Measurement of Heavy Metals Using ICP-MS Application
This application note describes a robust analytical method for the determination of heavy metals and substances regulated by the EU Restriction of Hazardous Substances/End of Life Vehicles (RoHS/ELV) directives. Based on the experience of Nihon Environmental Services Co., Ltd in Japan using an Agilent Technologies 7500c ORS (Octopole Reaction System) collision/reaction cell ICP-MS, this note su...
متن کاملSemiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode
Rapid semiquantitative elemental screening of pharmaceutical samples is described using the Agilent 7500ce ICP-MS with helium as collision gas. Polyatomic interferences are eliminated in the He pressurized cell, enabling rapid elemental profiling of pharmaceutical tablets for screening purposes. Elements which normally suffer from polyatomic interferences have been included in the semiquantitat...
متن کاملDirect Elemental Analysis of Gasoline by Agilent 7500ce ORS ICP-MS Application
The Agilent 7500ce ICP-MS is shown to be an excellent analytical tool for the elemental analysis of gasoline. It offers the best detection limits of any multi-element technique and a full elemental analysis suite that takes less than 5 minutes to acquire. The Octopole Reaction System (ORS), integral to the 7500ce, ensures that a single method and set of conditions can be used to directly determ...
متن کاملSupercritical Fluid Extraction of Metals Using Bis(2,4,4-trimethyl-pentyl)monothiophosphinic Acid as Chelating Agent for Subsequent ICP-MS Analyses of Mercury, Cadmium and Lead in Sediment
This study investigated and validated the optimum conditions for a supercritical fluid extraction (SFE) method using bis(2,4,4-trimethylpentyl)monothiophosphinic acid as a chelating ligand, for subsequent ICP-MS analyses of mercury, cadmium and lead in sediment. Several combinations of pressure, temperature and modifier (P-T-M) conditions for SFE were tested and an appropriate collection solven...