Validation of the Diesse Mini-Ves erythrocyte sedimentation rate (ESR) analyzer using the Westergren ESR method in patients with systemic inflammatory conditions.

نویسندگان

  • Marc R Happe
  • Daniel F Battafarano
  • David P Dooley
  • Thomas A Rennie
  • Frederick T Murphy
  • Thomas J Casey
  • John A Ward
چکیده

The Diesse Mini-Ves (DMV) erythrocyte sedimentation rate (ESR) analyzer was designed to improve efficiency in determining the ESR. We compared the Westergren ESR method with the 4-sample DMV ESR analyzer for performance and clinical correlation. This prospective observational study, conducted at a 450-bed tertiary medical center, evaluated 291 paired samples from subjects with various systemic inflammatory conditions. Linear regression analysis revealed a statistically significant correlation between the 2 methods. Satisfactory precision of the DMV analyzer was obtained for high and mid-range ESR values. The 4-sample DMV ESR analyzer was precise and comparable in results to the Westergren ESR method. This DMV ESR analyzer is now used at our medical center based on quality control improvements that include a faster, safer, and more standardized ESR method. Hospital or office-based clinical laboratories should consider using the 4-sample DMV ESR analyzer in place of the Westergren method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of Diesse Ves-matic automated system for measuring erythrocyte sedimentation rate.

Measurement of the erythrocyte sedimentation rate (ESR) using a closed tube system reduces the biohazard risk to laboratory staff. The Diesse Ves-matic system offers manual or vacuum collection of blood into plastic tubes, automated mixing of the sample, and automated reading of the end point after 20 minutes of sedimentation. This system was compared with the 1977 Westergren ESR method of the ...

متن کامل

Evaluation of the Ves-Matic Cube 200 erythrocyte sedimentation method: comparison with Westergren-based methods.

The erythrocyte sedimentation rate (ESR) is still a widely used parameter for acute phase inflammation. Recently, new methods based on direct undiluted measurement of ESR in a standard EDTA tube have been developed. We evaluated the analytic performance of one of these new methods, the Ves-Matic Cube 200 (Diesse Diagnostica Senese, Siena, Italy), and compared it with several established Westerg...

متن کامل

Internal Quality Control for Erythrocyte Sedimentation Rate Using Patient Blood Sample

The erythrocyte sedimentation rate (ESR) test continues to be used as an important tool in many laboratories; either manually or as automated method. When automation is applied in clinical laboratories, it is imperative that reliable methods for calibration and appropriate control materials are available for monitoring the accuracy and precision of the results generated. But when the commercial...

متن کامل

Evaluation of Ves-Matic Cube 200--an automated system for the measurement of the erythrocyte sedimentation rate.

Ves-Matic Cube 200 is fully automated analyzer that performs erythrocyte sedimentation rate (ESR) measurement using the standard ethylenediaminetetraacetic acid blood sample tube, thus markedly reducing the analytical time and avoiding the need for an extra blood sample. The aim of this study was to assess the automatic Ves-Matic Cube 200 system for the measurement of ESR in comparison with the...

متن کامل

Aberration of Erythrocyte Sedimentation Rate by Zinc Nanoparticles

Dear Editor, erythrocyte sedimentation rate (ESR) is a useful basic clinical pathology laboratory investigation. It can be helpful in diagnosis and follow-up of several diseases. At present, a new automated method with proven reliability is available for ESR test  (1). Here, the authors report on observation on a laboratory experiment to test the effect of zinc nanoparticles on ESR results. The...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • American journal of clinical pathology

دوره 118 1  شماره 

صفحات  -

تاریخ انتشار 2002