A measurement technique to determine the calibration accuracy of an electromagnetic tracking system to radiation isocenter.

نویسندگان

  • Dale W Litzenberg
  • Ian Gallagher
  • Kathryn J Masi
  • Choonik Lee
  • Joann I Prisciandaro
  • Daniel A Hamstra
  • Timothy Ritter
  • Kwok L Lam
چکیده

PURPOSE To present and characterize a measurement technique to quantify the calibration accuracy of an electromagnetic tracking system to radiation isocenter. METHODS This technique was developed as a quality assurance method for electromagnetic tracking systems used in a multi-institutional clinical hypofractionated prostate study. In this technique, the electromagnetic tracking system is calibrated to isocenter with the manufacturers recommended technique, using laser-based alignment. A test patient is created with a transponder at isocenter whose position is measured electromagnetically. Four portal images of the transponder are taken with collimator rotations of 45° 135°, 225°, and 315°, at each of four gantry angles (0°, 90°, 180°, 270°) using a 3×6 cm2 radiation field. In each image, the center of the copper-wrapped iron core of the transponder is determined. All measurements are made relative to this transponder position to remove gantry and imager sag effects. For each of the 16 images, the 50% collimation edges are identified and used to find a ray representing the rotational axis of each collimation edge. The 16 collimator rotation rays from four gantry angles pass through and bound the radiation isocenter volume. The center of the bounded region, relative to the transponder, is calculated and then transformed to tracking system coordinates using the transponder position, allowing the tracking system's calibration offset from radiation isocenter to be found. All image analysis and calculations are automated with inhouse software for user-independent accuracy. Three different tracking systems at two different sites were evaluated for this study. RESULTS The magnitude of the calibration offset was always less than the manufacturer's stated accuracy of 0.2 cm using their standard clinical calibration procedure, and ranged from 0.014 to 0.175 cm. On three systems in clinical use, the magnitude of the offset was found to be 0.053±0.036, 0.121±0.023, and 0.093±0.013 cm. CONCLUSIONS The method presented here provides an independent technique to verify the calibration of an electromagnetic tracking system to radiation isocenter. The calibration accuracy of the system was better than the 0.2 cm accuracy stated by the manufacturer. However, it should not be assumed to be zero, especially for stereotactic radiation therapy treatments where planning target volume margins are very small.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method

Background: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Objective: A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...

متن کامل

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

متن کامل

Infrared Thermopile Temperature Measurement Technique in Microwave Heating Systems

Temperature measurement in microwave systems is essential for thermally driven processes, namely, catalytic reactions and ceramic sintering. Although, the application of direct thermometry methods, namely, thermocouples, have been commonly articulated in the available literature, however, contacted temperature measurement mechanisms have aroused concerns associated with the disruption of the el...

متن کامل

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...

متن کامل

SU-E-T-24: Development and Implementation of an Automated Algorithm to Determine Radiation Isocenter, Radiation vs. Light Field Coincidence, and Analyze Strip Tests.

PURPOSE The aim of this study was to develop a phantom and analysis software that could be used to quickly and accurately determine the location of radiation isocenter using the Electronic Portal Imaging Device (EPID). The phantom could then be used as a static reference point for performing other tests including: radiation vs. light field coincidence, MLC and Jaw strip tests, and Varian Optica...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 40 8  شماره 

صفحات  -

تاریخ انتشار 2013