Development and performance of a fully automated method for assay of C-reactive protein in the aca discrete clinical analyzer.
نویسندگان
چکیده
A quantitative immunoassay for C-reactive protein (CRP) has been developed for use in the Du Pont aca discrete clinical analyzer. Particle-enhanced turbidimetric immunoassay (PE-TIA) technology is used. The method has a CV of less than 10% in the range 2 to 120 mg/L. Neither hemolyzed samples (Hb less than 5 g/L), icteric samples (total bilirubin less than 300 mg/L), lipemic samples (triglyceride less than 15 g/L), nor some commonly used drugs interfere. Dithioerythritol is used to eliminate interference from rheumatoid factor. Good correlation was seen when the Du Pont CRP method was compared with the Beckman ICS, Syva EMIT, TDx, and Behring methods for CRP. The normal reference interval is 0 to 9 mg/L. The method, which is fully automated, is fast, requires only a few microliters of serum, and is well suited to emergency-room requirements.
منابع مشابه
Improvement of anticomplementary activity assay for the quality control of immunoglobulin product: an Iranian pilot scale study
Abstract Background and Objectives One of the side effects of immunoglobulin administration is the activation of the complement system related to the high concentration of aggregates/large polymers in the production process. Measuring anticomplementary activity (ACA) is important in pharmaceutical quality control and is a criterion of spontaneous activation of complement system in immunoglobuli...
متن کاملDevelopment and characterization of an automated assay of effective heparin activity in plasma.
We describe a fully automated assay for determining effective heparin activity in plasma, based on heparin-catalyzed inhibition of Factor Xa (EC 3.4.21.6) by antithrombin III (AT III). Residual Factor Xa is determined kinetically by the Du Pont aca discrete clinical analyzer with a chromogenic substrate and is inversely related to heparin activity. Because the test plasma is the sole source of ...
متن کاملProcalcitonin: A Reliable Marker for the Diagnosis of Neonatal Sepsis
Objective(s) In the last few years, serum procalcitonin has been proposed as an early marker of infections in neonates, with varying results. In this study, we aimed to investigate the value of procalcitonin, and C- reactive protein in establishing the diagnosis of neonatal sepsis. Materials and Methods Blood samples were collected at admission from 69 neonates with suspected infection (admi...
متن کاملA quantitative automated immunoassay for fibrinogen/fibrin degradation products.
We describe a prototype quantitative automated assay for fibrin and fibrinogen degradation products, a particle-enhanced turbidimetric inhibition immunoassay (PETINIA) in the Du Pont aca discrete clinical analyzer. This assay involves a latex particle reagent with covalently bound fibrinogen and a polyclonal antiserum raised in rabbits against human fibrinogen. A special secondary sample-collec...
متن کاملPerformance characteristics of three automated immunoassays for thyroid hormones
Background: Since the introduction of the first radioimmunoassay, several improvements have been made in the design of immunoassays such as method of antibody production, labeling, automation and detection technology. We performed an analytical evaluation of the new electrochemiluminescent immunoassay (ECLIA) for serum TSH, FT4 and T3 in the Elecsys 2010 immunoassay system and compared the resu...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Clinical chemistry
دوره 34 8 شماره
صفحات -
تاریخ انتشار 1988