A comparison of leak compensation during pediatric noninvasive ventilation: a lung model study.

نویسندگان

  • Jun Oto
  • Christopher T Chenelle
  • Andrew D Marchese
  • Robert M Kacmarek
چکیده

BACKGROUND Ventilators used for noninvasive ventilation (NIV) must be able to synchronize in the presence of system leaks. We compared the ability of 7 ICU ventilators and 3 dedicated NIV ventilators to compensate for leaks during pediatric NIV. METHODS Using a lung simulator, we compared the Maquet Servo-i, Dräger V500, Dräger Carina, Covidien PB840, Respironics V60, Respironics Vision, GE Healthcare/Engström Carestation, CareFusion Avea, Hamilton C3, and Hamilton G5 during increasing (n = 6) and decreasing leaks (n = 6). With a lung simulator we tested 4 leak levels (baseline [BL] 2-3 L/min, L1 5-6 L/min, L2 9-10 L/min, and L3 19-20 L/min); 3 patient weights (10, 20, and 30 kg); and 3 lung mechanics scenarios (normal, obstructive, and restrictive). The ventilator settings were NIV mode, pressure support of 10 cm H2O, and PEEP of 5 cm H2O. The synchronization rate (synchronized cycles/total simulated respirations) was recorded for each ventilator and each leak scenario. Synchronization was defined as triggering without auto-triggering, miss-triggering, delayed cycling, or premature cycling. RESULTS The mean synchronization rate across all ventilators was 68 ± 27% (range 23-96%) and marked differences existed between the ventilators (P < .001). Significant differences in synchronization rate were observed between the 10-kg model (mean 57 ± 30%, range 17-93%), the 20-kg model (69 ± 30%, 25-98%), and the 30-kg models (77 ± 22%, 28-97%) (P < .001). The synchronization rate for the obstructive model (60 ± 30%, 9-94%) was significantly different from the normal model (71 ± 29%, 18-98%) and the restrictive model (72 ± 28%, 23-98%) (P < .001). The PB840 and the C3 had synchronization rates over 90% overall across all body weights, all lung mechanic profiles, and all leak levels. CONCLUSIONS Leak compensation in NIV for pediatric use can partially compensate for leaks, but varies widely among ventilators, patient weights, and lung mechanics.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A comparison of leak compensation in acute care ventilators during noninvasive and invasive ventilation: a lung model study.

BACKGROUND Although leak compensation has been widely introduced to acute care ventilators to improve patient-ventilator synchronization in the presence of system leaks, there are no data on these ventilators' ability to prevent triggering and cycling asynchrony. The goal of this study was to evaluate the ability of leak compensation in acute care ventilators during invasive and noninvasive ven...

متن کامل

Bronchodilator delivery during simulated pediatric noninvasive ventilation.

BACKGROUND Noninvasive ventilation (NIV) is usually applied using bi-level positive airway pressure devices, and many of these devices use a single-limb patient circuit with an integrated leak port to purge the circuit of exhaled carbon dioxide. Sometimes bronchodilator therapy is indicated in pediatric patients, but there have been no studies of the optimal nebulizer position, with respect to ...

متن کامل

Leak compensation in positive pressure ventilators: a lung model study.

Leak compensating abilities of six different positive pressure ventilators commonly used to deliver noninvasive positive pressure ventilation, including the bilevel positive airway pressure (BiPAP) S/T-D and Quantum (Respironics Inc, Murrysville, PA, USA), 335 and O'NYX (Mallinckrodt Inc, St Louis, MO, USA), PLV 102 (Respironics), and Siemens Servo 900C (Siemens Inc, Danvers, MA, USA). Using a ...

متن کامل

Comparative evaluation of LMA-Supreme and i-gel supraglottic airway devices with endotracheal intubation during surgical correction of traumatic orbital injuries

Aims. The aim of the study was to compare endotracheal intubation with i-gel and LMA (laryngeal mask airway) supreme supraglottic airway devices in orbital osteosynthesis surgery. Methods. 134 patients undergoing orbital plasty after fracture were enrolled in our study and were randomized into three groups: 43 in the endotracheal intubation group, 42 in the LMA-Supreme group, and 49 patients w...

متن کامل

Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation.

OBJECTIVES During noninvasive positive pressure ventilation (NPPV), the gas leak that commonly occurs around the mask can render NPPV ineffective. We evaluated the effects of gas leak on inspiratory trigger function during NPPV with bilevel pressure and ICU ventilators. In addition, we evaluated the effects of gas leak on fraction of inspired oxygen (Fio(2)) and humidification. METHODS Air le...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Respiratory care

دوره 59 2  شماره 

صفحات  -

تاریخ انتشار 2014