Airway obstruction by dissection of the inner layer of a reinforced endotracheal tube in a patient with Ludwig's angina: A case report
نویسندگان
چکیده
Intraoperative airway obstruction is perplexing to anesthesiologists because the patient may fall into danger rapidly. A 74-year-old woman underwent an emergency incision and drainage for a deep neck infection of dental origin. She was orally intubated with a 6. 0 mm internal diameter reinforced endotracheal tube by video laryngoscope using volatile induction and maintenance anesthesia (VIMA) with sevoflurane, fentanyl (100 µg), and succinylcholine (75 mg). During surgery, peak inspiratory pressure increased from 22 to 38 cmH2O and plateau pressure increased from 20 to 28 cmH2O. We maintained anesthesia because we were unable to access the airway, which was covered with surgical drapes, and tidal volume was delivered. At the end of surgery, we found a longitudinal fold inside the tube with a fiberoptic bronchoscope. The patient was reintubated with another tube and ventilation immediately improved. We recognized that the tube was obstructed due to dissection of the inner layer.
منابع مشابه
Partial airway obstruction caused by dissection of a reinforced endotracheal tube.
EDITOR: Ventilatory problems during surgery in the prone position may be a serious complication [1]. We report an incident where there was dissection of a reinforced endotracheal tube that led to its partial obstruction. This case shows an unexpected complication from reusing products intended for single use. A 62-yr-old female (weight 68 kg) who was scheduled for total laminectomy with posteri...
متن کاملImpending Complete Airway Obstruction from a Reinforced Orotracheal Tube: a Case Report.
Reinforced tubes are commonly used to minimize the opportunity of upper airway obstruction in patients at risk. There are a few reports of the airway obstruction resulted from kinked reinforced tubes. This report describes the obstruction of a reinforced tube in an adult patient who underwent tonsillectomy. Under general anesthesia; an armoured endotracheal tube was inserted into the trachea un...
متن کاملSafe Use of Flammable Endotracheal Tubes During Anesthesia for Laryngeal Laser Surgery-Report of 1024 Cases and a Brief Literature Review
Introduction: One of the major concerns in laryngeal laser surgery is the risk of airway fire. The introduction ofwrapped tubes and metal tubes has reduced the fire hazards. However, these tubes are expensive and do not provide convenient access to the surgical field. There are few laboratory studies addressing the resistance of polyvinylchloride tubes against ignition in the given circum...
متن کامل[Airway management in Ludwig's angina - a challenge: case report].
BACKGROUND Ludwig's angina (LA) is an infection of the submandibular space, first described by Wilhelm Frederick von Ludwig in 1836. It represents an entity difficult to manage due to the rapid progression and difficulty in maintaining airway patency, a major challenge in medical practice, resulting in asphyxia and death in 8-10% of patients. OBJECTIVE Describe a case of a patient with Ludwig...
متن کامل[Dissection of the wired endotracheal tube's lumen during general anesthesia: a case report].
OBJECTIVE The aim of this study is to report a case of a clinically significant obstruction during mechanical ventilation caused by the dissection of the wired endotracheal tube's lumen during general anesthesia in a pediatric patient. CASE REPORT A12-years old patient undergoing general anesthesia for open appendectomy was intubated with a wired endotracheal tube and difficult removal of the...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 17 شماره
صفحات -
تاریخ انتشار 2017