Dosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.

نویسندگان

  • Mark J Rivard
  • Christopher S Melhus
  • David E Wazer
  • Raymond J Bricault
چکیده

PURPOSE The AccuBoost brachytherapy system applies HDR 192Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR 192Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use. METHODS Dosimetric measurements were performed in a polystyrene phantom, while simulations estimated dose in air, liquid water, polystyrene and ICRU 44 breast tissue. Dose distribution characterization of the 4-8 cm diameter collimators was performed using radiochromic EBT film and air ionization chambers. RESULTS The central axis dose falloff was steeper for the 4 cm diameter applicator in comparison to the 8 cm diameter applicator, with surface to 3 cm depth-dose ratios of 3.65 and 2.44, respectively. These ratios did not considerably change when varying the phantom composition from breast tissue to polystyrene, phantom thickness from 4 to 8 cm, or phantom radius from 8 to 15 cm. Dose distributions on the central axis were fitted to sixth-order polynomials for clinical use in a hand calculation spreadsheet (i.e., nomogram). Dose uniformity within the useful applicator apertures decreased as depth-dose increased. CONCLUSIONS Monte Carlo benchmarking simulations of the HDR 192Ir source using the MCNP5 radiation transport code indicated agreement within 1% of the published results over the radial/angular region of interest. Changes in phantom size and radius did not cause noteworthy changes in the central axis depth-dose. Polynomial fit depth-dose curves provide a simple and accurate basis for a nomogram.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost

Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...

متن کامل

The Effect of Breast Phantom Material on the Dose Distribution in AccuBoost Brachytherapy

Introduction: Long-term teletherapy program is not suitable for old and working patients and those living in areas where little access to primary health care is available. Accelerated partial breast irradiation (APBI) using high dose rate (HDR) brachytherapy is an appropriate alternative for these patients due to its limited number of fractions. The AccuBoost is a system for delivering APBI. Th...

متن کامل

Dosimetric optimization of a conical breast brachytherapy applicator for improved skin dose sparing.

PURPOSE Both the AccuBoost D-shaped and round applicators have been dosimetrically characterized and clinically used to treat patients with breast cancer. While the round applicators provide conformal dose coverage, under certain clinical circumstances the breast skin dose may be higher than preferred. The purpose of this study was to modify the round applicators to minimize skin dose while not...

متن کامل

Evaluation of BEBIG HDR 60Co system for non-invasive image-guided breast brachytherapy

PURPOSE HDR (60)Co system has recently been developed and utilized for brachytherapy in many countries outside of the U.S. as an alternative to (192)Ir. In addition, the AccuBoost(®) technique has been demonstrated to be a successful non-invasive image-guided breast brachytherapy treatment option. The goal of this project is to evaluate the possibility of utilizing the BEBIG HDR (60)Co system f...

متن کامل

Evaluation of dose distribution and dose gradient in brachytherapy cylindrical applicators using a dedicated Phantom for Iridium-192 and Cobalt-60 HDR sources

Introduction: A study was performed to evaluate radiation dose distribution and dose gradient around cylindrical applicators for high-dose-rate (HDR) brachytherapy systems with 192Ir, 60Co brachytherapy source applied for rectal and vaginal cancers treatments.   Materials and Methods: Two additional computed tomography (CT) based plans were generated using a ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 36 11  شماره 

صفحات  -

تاریخ انتشار 2009