International experience with secundum atrial septal defect occlusion by the buttoned device.

نویسندگان

  • P S Rao
  • E B Sideris
  • G Hausdorf
  • C Rey
  • T R Lloyd
  • R H Beekman
  • A M Worms
  • F Bourlon
  • E Onorato
  • M Khalilullah
چکیده

Several devices are available for transcatheter occlusion of atrial septal defect. This report describes the international experience with the buttoned device. During a 4.5-year period ending in February 1993, 180 transcatheter atrial septal defect occlusions were performed with the buttoned device. Patient age varied between 0.6 and 76 years and stretched atrial defect diameter between 5 and 25 mm. The defects were closed with 25 to 50 mm devices delivered through 8F (148 patients) or 9F (32 patients) sheaths. Twelve patients were adults whose defects were closed to prevent recurrence of cerebrovascular accidents caused by presumed paradoxic embolism. In the remaining patients the atrial defect was closed to treat the left-to-right shunt. The atrial septal defects were effectively occluded as demonstrated by (1) decrease in pulmonary-to-systemic flow ratio from 2.1 +/- 0.6 (mean +/- SD) to 1.05 +/- 0.1 (p < 0.01) by oximetry; (2) normalized S2 and disappearance of the diastolic murmur by auscultation; and (3) improvement in right ventricular volume overloading by echocardiogram. However, trivial to small shunts could be detected by color Doppler studies in 76 (45%) of 168 patients in whom such data are available. Complications included unbuttoning in 13 and whole-device embolization in 1. All patients remained stable, and retrieval of the device and surgical closure of the atrial septal defect were accomplished in 10 patients. Transcatheter retrieval was used in the remaining 4 patients. The incidence of unbuttoning, a major complication of the procedure, appeared to decrease with the increasing experience of the investigators and with device modification (third-generation). The follow-up duration varied between 1 month and 4 years. Six patients required surgery during the follow-up period. In the remaining patients (n = 160), clinical examination did not reveal signs of atrial shunts. Color Doppler studies revealed either complete disappearance of the previously demonstrated shunts or further diminution of their size. The results indicate that transcatheter occlusion of the atrial septal defects with buttoned devices is feasible, relatively safe, and effective, and it appears to be a viable alternative to surgery for some patients with secundum atrial septal defect. Complications are infrequent and should improve with experience.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transcatheter closure of residual atrial septal defect following implantation of buttoned device.

We report a case in which residual shunting after a buttoned device occlusion of atrial septal defect (ASD) was eliminated by transcatheter retrieval of a portion of the device, followed by implantation of a second device. This method may be helpful for those patients with residual ASDs who decline surgical device retrieval and defect closure.

متن کامل

Integrated Percutaneous Atrial Septal Defect Occlusion and Pulmonary Balloon Valvuloplasty

Introduction: Atrial Septal Defect (ASD) is one of the most common congenital heart diseases during childhood. Today, ASD closure is done by occlutech device via cardiac catheterization. ASD repair with transcatheter technique has shown high closure rate. However, coexistence of severe pulmonary stenosis and large defect in atrial septum is rare. When these two problems coexist, the result is i...

متن کامل

Transvenous atrial septal defect occlusion in piglets with a "buttoned" double-disk device.

The feasibility and safety of transvenous closure of atrial septal defects by a new device was tested in 20 piglets, in which atrial septal defects were created by foramen ovale dilatation with angioplasty balloons. The device was small enough to be introduced in a 7F or 8F sheath, and it measured 20-25 mm. It has no hooks and consists of a foam occluder from the left atrium and a counter-occlu...

متن کامل

Late migration of a Sideris buttoned device for occlusion of atrial septal defect.

We describe a 17-year-old patient with an atrial septal defect who underwent device closure with a second generation Sideris buttoned device at 4 years of age. She presented 13 years after the procedure with the acute onset of chest discomfort, at which time a chest radiograph showed migration of the wire of the right atrial counter-occluder to the lateral aspect of the right atrial wall, with ...

متن کامل

Doppler forward flow profiles of St. Jude Medical prosthetic valves in pediatric patients.

tion versus transcatheter closure of ASD before comparisons can he made. The length of stay for patients undergoing transcatheter closure is reported to he 2 to 3 days.1-5y9 In our study, the most recent average length of stay was 6.1 days, hut since this study has ended, the length of stay for ASD operation is ~4 days. Recent reports on transcatheter closure devices do not discuss costs, inclu...

متن کامل