Inaccuracy of tidal volume delivered by home mechanical ventilators.
نویسندگان
چکیده
Ideally, the inspired (tidal) volume (V(T)) provided by a volume-controlled ventilation device should not change when the pressure imposed on the ventilator varies. A bench study evaluation of V(T) versus pressure was performed on 10 commercially available devices. The difference between the desired V(T) and the observed V(T) reached 100 mL for some devices when inspiratory resistance was at its lowest, rising to 150 mL when inspiratory resistance was increased to obtain peak airway pressure of 60 cmH2O. The present data indicate that some home ventilators are inaccurate in delivering the preset tidal volume when the pressure imposed on the ventilator is increased to simulate high airway resistance.
منابع مشابه
Near-infrared spectroscopy with vascular occlusion test may not be the adequate tool to explore microcirculation in pulmonary arterial hypertension.
essentials of respiratory care, 4th edition. St. Louis:Mosby Elsevier; 2005:103. 13. L’Her E, Roy A. Bench tests of simple, handy ventilators for pandemics: performance, autonomy, and ergonomy. Respir Care 2011;56(6):751-760. 14. Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC. Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventila...
متن کاملEffect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.
BACKGROUND New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE To assess the reliability of the VT provided by 5 ventilators in a bench test. METHODS Five commercial ventilators from 4 different manufacturers were tested in pr...
متن کاملRESPIRATION AND THE AIRWAY Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study
Editor’s key points † This study investigated the accuracy of delivered tidal volume in four new ventilators under different conditions. † Set and delivered tidal volumes differed when compliance and resistance were changed, although the effect of fresh gas flow was less. † Some of the new ventilators tested were more accurate than others. † The clinical significance of these findings is uncert...
متن کاملActual performance of mechanical ventilators in ICU: a multicentric quality control study
Even if the performance of a given ventilator has been evaluated in the laboratory under very well controlled conditions, inappropriate maintenance and lack of long-term stability and accuracy of the ventilator sensors may lead to ventilation errors in actual clinical practice. The aim of this study was to evaluate the actual performances of ventilators during clinical routines. A resistance (7...
متن کاملTemperature of gas delivered from ventilators
BACKGROUND Although heated humidifiers (HHs) are the most efficient humidifying device for mechanical ventilation, some HHs do not provide sufficient humidification when the inlet temperature to the water chamber is high. Because portable and home-care ventilators use turbines, blowers, pistons, or compressors to inhale in ambient air, they may have higher gas temperature than ventilators with ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The European respiratory journal
دوره 15 2 شماره
صفحات -
تاریخ انتشار 2000