A detailed measurement of distortion product otoacoustic emission (DPOAE) fine structure was used to extract estimates of the two major components believed to contribute to the overall DPOAE level in the ear canal. A fixed-ratio paradigm was used to record DPOAE fine structure from three normal-hearing ears over a range of 400 Hz for 12 different stimulus-frequency ratios between 1.053 and 1.36 and stimulus levels between 45 and 75 dB SPL. Inverse Fourier transforms of the amplitude and phase data were filtered to extract the early component from the generator region of maximum stimulus overlap and the later component reflected from the characteristic frequency region of the DPOAE. After filtering, the data were returned to the frequency domain to evaluate the impact of the stimulus-frequency ratio and stimulus level on the relative levels of the components. Although there were significant differences between data from different ears some consistent patterns could be detected. The component from the overlap region of the stimulus tones exhibits a bandpass shape, with the maximum occurring at a ratio of 1.2. The mean data from the DPOAE characteristic frequency region also exhibits a bandpass shape but is less sharply tuned and exhibits greater variety across ears and stimulus levels. The component from the DPOAE characteristic frequency region is dominant at ratios narrower than approximately 1.1 (the transition varies between ears). The relative levels of the two components are highly variable at ratios greater than 1.3 and highly dependent on the stimulus level. The reflection component is larger at all ratios at the lowest stimulus level tested (45/45 dB SPL). We discuss the factors shaping DPOAE-component behavior and some cursory implications for the choice of stimulus parameters to be used in clinical protocols.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Characteristics of the 2f(1)-f(2) distortion product otoacoustic emission in a normal hearing population.

Distortion-product otoacoustic emission (DPOAE) fine structure and component characteristics are reported between 0.75 and 16 kHz in 356 clinically normal hearing human subjects ages 10 to 65 yr. Stimulus tones at 55/40, 65/55, and 75/75 dB SPL were delivered using custom designed drivers and a calibration method that compensated for the depth of insertion of the otoacoustic emission (OAE) prob...

متن کامل

Evaluation of Distortion and Transient Evoked Otoacoustic Emission in Tinnitus Patients with Normal Hearing

Introduction: Tinnitus is a perception of sound without external source. The exact etiology of tinnitus is not fully understood, although some researchers believe that the condition usually starts in the cochlea. The aim of this study was to determine the potential contribution of outer hair cell dysfunction to chronic tinnitus, by application of Distortion-Product Evoked Otoacoustic Emission (...

متن کامل

Mechanisms and Applications

product (DP), distortion-product “audiogram” (DP-gram), distortion-product OAE (DPOAE), fine structure, fast Fourier transform (FFT), input/output (I/O), inverse FFT (IFFT), magnitude and phase, medial olivocochlear (MOC), otoacoustic emissions (OAE), outer hair cell (OHC), polar plot, reflection, signal-to-noise ratio (SNR), spontaneous OAE (SOAE), stimulus-frequency OAE (SFOAE), transient-evo...

متن کامل

Variation of distortion product otoacoustic emissions with furosemide injection.

Cochlear function was monitored in adult gerbils using distortion product otoacoustic emissions (DPOAE) during intraperitoneal injection of furosemide. All stimulus parameters were varied independently over a wide range, the stimulus frequencies f1 and f2 from 1 to 16 kHz, and the stimulus levels L1 and L2 from 20 to 80 dB SPL. The observed emissions at 2f1-f2 and 3f1-2f2 could be considered to...

متن کامل

Input-output functions of the nonlinear-distortion component of distortion-product otoacoustic emissions in normal and hearing-impaired human earsa)

Distortion-product otoacoustic emissions (DPOAEs) arise in the cochlea in response to two tones with frequencies f1 and f2 and mainly consist of two components, a nonlinear-distortion and a coherent-reflection component. Wave interference between these components limits the accuracy of DPOAEs when evaluating the function of the cochlea with conventional continuous stimulus tones. Here, DPOAE co...