Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements

نویسندگان

  • J. Saarelainen
  • M. Hakulinen
  • T. Rikkonen
  • H. Kröger
  • M. Tuppurainen
  • H. Koivumaa-Honkanen
  • R. Honkanen
  • M. Hujo
  • J. S. Jurvelin
چکیده

In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21-72 years) underwent anteroposterior lumbar spine L2-L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r = 0.962-0.995; p < 0.0001). The average iDXA BMD values were 1.5%, 0.5%, and 0.9% higher than those of Prodigy for lumbar spine (L2-L4) (p < 0.0001), femoral neck (p = 0.048), and total hip (p < 0.0001), respectively. Total body BMD values measured with the iDXA were -1.3% lower (p < 0.0001) than those measured with the Prodigy. For total body, lumbar spine, and femoral neck, the BMD differences as measured with these two devices were independent of subject height and weight. Linear correction equations were developed to ensure comparability of BMD measurements obtained with both DXA scanners. Importantly, use of equations from previous studies would have increased the discrepancy between these particular DXA scanners, especially at hip and at spine.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cross calibration of two dual-energy X-ray densitometers and comparison of visceral adipose tissue measurements by iDXA and MRI

OBJECTIVE Compare total percentage body fat (pfat) measurements between two densitometers and visceral adipose tissue (VAT) estimates between iDXA and magnetic resonance imaging (MRI) from the same defined abdominal region. METHODS Participants [N = 93 (50 men, 43 women), BMI: 19.1-57.6 kg/m2 ] underwent dual-energy X-ray absorptiometry (DXA) scans on two DXA systems (GE Healthcare Lunar iDXA...

متن کامل

Dual energy x-ray absorptiometry-based assessment of male patients using standardized bone density values and a national reference database.

Dual energy X-ray absorptiometry (DXA) measurements from different manufacturers provide different bone mineral density (BMD) values and derived T-scores and Z-scores. These differences result partly from technical differences in the algorithms for the determination of bone mineral content and bone area and partly from the use of different manufacturer-derived reference databases. The present s...

متن کامل

Bone mineral density in elite rowers

Methods Subjects (n=125) were internationally competitive. Between 2011-2014 BMD was taken at the lumbar spine (L1-L4) and left femur, by Dual-energy X-ray absorptiometry (DXA, Lunar Prodigy, GE Healthcare), using the same scanner, and a qualified technician. Ethics was approved by the Australian Institute of Sport Human Ethics committee. Subjects gave prior written informed consent. Descriptiv...

متن کامل

Comparison of Bone Mineral Density Measured by Dual X-ray, Axial Dual-energy Photon X-ray Absorptiometry and Laser Absorptiometry of Calcaneus

The objective of this study was to validate the use of bone mineral measurements of the calcaneus bone by dual X-ray and laser (DXL) in a cross-sectional study carried out in an osteoporosis clinic. Measurements of bone mineral density (BMD) at proximal femur and spine were obtained by dual-energy photon X-ray absorptiometry (DEXA). Osteoporosis was defined by a DEXA T-score <-2.5 at the femora...

متن کامل

Amalgamated Reference Data for Size-Adjusted Bone Densitometry Measurements in 3598 Children and Young Adults-the ALPHABET Study.

The increasing use of dual-energy X-ray absorptiometry (DXA) in children has led to the need for robust reference data for interpretation of scans in daily clinical practice. Such data need to be representative of the population being studied and be "future-proofed" to software and hardware upgrades. The aim was to combine all available pediatric DXA reference data from seven UK centers to crea...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 2016  شماره 

صفحات  -

تاریخ انتشار 2016