Selectivity Enhancement for Chloride Ion by In(III)-Porphyrin-Based Polymeric Membrane Electrode Operated in Pulsed Chronopotentiometric Mode.
نویسندگان
چکیده
A robust selectivity enhancement of an In(III)-porphyrin ionophore-based chloride-selective electrode under pulsed chronopotentiometric measurement mode that enables the detection of chloride ions in the presence of a normally interfering concentration of salicylate ions is described. This enhancement is achieved by the rapid depletion of the surface concentration of the more dilute lipophilic anion during an initial anodic current pulse period due to extraction of this preferred anion into the membrane phase. Measurement of chloride with a detection limit of 8 mM and near Nernstian response slope in the presence of 1 mM salicylate is possible using the pulstrode method.
منابع مشابه
Polymeric Membrane Sensor for Potentiometric Determination of Tin(II) Ions
Tin(II)-ion-selective electrode consisting of dibenzo-18-crown-6 (DB18C6) as an ionophore inPVC matrix was developed. The influences of membrane composition, selectivity, response time,reversibility and temperature on the electrode performance were investigated. The electrodeshowed a near Nernstian response for Sn2+ ions over a concentration range from 1.0× 10-6 – 1.0× 10-2 M with a slope of 27...
متن کاملReversible detection of proteases and their inhibitors by a pulsed chronopotentiometric polyion-sensitive electrode.
Polymer membrane electrodes operated by pulsed chronopotentiometry have recently been introduced to replace traditional ion-selective electrodes for a number of applications. While ion-selective electrodes for the polycation protamine have been reported, for instance, a pulsed chronopotentiometric readout mode (called here pulstrode) provides improved stability and reproducibility while exhibit...
متن کاملFe (II) Ion-Selective Membrane Electrode based onTetra-Phenyl Porphyrin in PVC Matrix
Fe2+ ion-selective membrane sensor has been fabricated from polyvinyl chloride (PVC) matrixmembrane containing neutral carrier tetra phenyl porphyrin (TPP) ionophore. The addition ofsodium tetraphenylborate (NaTPB) and the plasticizer DBP has been found to substantiallyimprove the performance of the sensor. The best performance was obtained with the sensor havinga membrane of composition of TPP...
متن کاملAluminum(III) Porphyrins as Ionophores for Fluoride Selective Polymeric Membrane Electrodes
Aluminum(III) porphyrins are examined as potential fluoride selective ionophores in polymeric membrane type ionselective electrodes. Membranes formulated with Al(III) tetraphenyl (TPP) or octaethyl (OEP) porphyrins are shown to exhibit enhanced potentiometric selectivity for fluoride over more lipophilic anions, including perchlorate and thiocyanate. However, such membrane electrodes display un...
متن کاملApplication of a dioxime-PVC electrode to potentiometric studies of Cr(III) ion
A PVC membrane Cr(III) ion selective electrode has been constructed using 2H-1,4-benzothioazine-2,3(4H)dione dioxime (BTD) as membrane carrier. The influence of membrane composition on the electroderesponse was studied. The electrode exhibits a Nernstian response over a Cr(III) concentration range of 1.0 x10 6 to 1.0>< 10-1 M (r = 0.99) with a slope of 19.5 ± 0.5 mV per decade of concentration,...