Assessment of left ventricular volumes using simplified 3-D echocardiography and computed tomography – a phantom and clinical study
نویسندگان
چکیده
OBJECTIVES To compare the accuracy of simplified 3-dimensional (3-D) echocardiography vs. multi-slice computed tomography (MSCT) software for the quantification of left ventricular (LV) volumes. DESIGN Three-D echocardiography (3-planes approach) and MSCT-CardIQ software were calibrated by measuring known volumes of 10 phantoms designed to closely mimic blood-endocardium interface. Subsequently, LV volumes were measured with both the methods in 9 patients referred routinely for coronary angiography and the agreement between the measurements was evaluated. RESULTS Simplified 3D-echocardiography provided higher degree of agreement between the measured and true phantom volumes (mean difference 0 +/- 1 ml, variation range +4 to -4 ml) than MSCT software (mean difference 6 +/- 5 ml; variation range +22 to -10 ml). The agreement between LV measurements in the patients was considerably poorer, with significantly larger volumes produced by MSCT (mean difference -23 +/- 40 ml, variation between +93 and -138 ml). CONCLUSION Simplified 3-D echocardiography provides more accurate assessment of phantom volumes than MSCT-CardIQ software. The discrepancy between the results of LV measurements with the two methods is even greater and does not warrant their interchangeable diagnostic use.
منابع مشابه
High myocardial 18F-FDG uptake after chemotherapy in the presence of left ventricular dysfunction as well as 3-month later with no left ventricular dysfunction
Diffusely increased glucose metabolic activity in the right and left ventricles using 2-deoxy-2-(F-18) fluoro-D-glucose positron emission tomography/computed tomography (18F-FDGPET/CT) and global left ventricular hypokinesia at echocardiography can be evident in acute myocarditis. But, there has been no case report that the ventricular 18F-FDGuptake remains unchanged even ...
متن کاملPrediction of Changes in Left Ventricular Ejection Fraction after Off-Pump Coronary Artery Bypass Grafting Surgery by Myocardial Perfusion Single-Photon Emission Computed Tomography
Introduction: Left ventricular ejection fraction (LVEF) is considered to be the single most important prognostic factor in patients with previous myocardial infarction. LVEF is not improved in all patients after coronary artery bypass grafting (CABG). This study aimed to assess the possibility of prediction of LVEF changes after CABG using myocardial perfusion gated signle photon emission compu...
متن کاملAssessment of global and regional left ventricular function and volumes with 64-slice multi-slice computed tomography: a comparison with 2D echocardiography
Introduction: In patients with coronary artery disease (CAD), LV function and volumes are important parameters for long-term prognosis. 64-slice MSCT allows non-invasive assessment of the coronary arteries, but the accuracy of 64-slice MSCT for assessment of LV volumes and function is unknown. Methods: A head-to-head comparison between 64-slice MSCT and 2D echocardiography was performed in 40 p...
متن کاملRapid three-dimensional echocardiography : clinically feasible alternative for precise and accurate measurement of left ventricular volumes.
BACKGROUND Clinical applicability of conventional ultrasonographic systems using mechanical adapters for 3D echocardiographic imaging has been limited by long acquisition and processing times. We developed a rapid (6-s) acquisition technique that collects apical tomograms using a continuously internally rotating transthoracic transducer. This study was performed to examine the clinical feasibil...
متن کاملAssessment of global left ventricular function and volumes with 320-row multidetector computed tomography: A comparison with 2D-echocardiography
BACKGROUND Multidetector computed tomography (MDCT) has been demonstrated as a feasible imaging modality for noninvasive assessment of coronary artery disease and left ventricular (LV) function. Recently, 320-row systems have become available with 16 cm anatomical coverage allowing image acquisition of the entire heart within a single heartbeat. The purpose of this study was to evaluate the acc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Cardiovascular Ultrasound
دوره 6 شماره
صفحات -
تاریخ انتشار 2008