Cerecyte versus platinum coils in the treatment of intracranial aneurysms: packing attenuation and clinical and angiographic midterm results.

نویسندگان

  • I Linfante
  • M J DeLeo
  • M J Gounis
  • C S Brooks
  • A K Wakhloo
چکیده

BACKGROUND AND PURPOSE As modifications are made to coils, monitoring the safety profile, ability to achieve high packing attenuation, and durability of occlusion as compared to the standard bare platinum coils is of paramount importance. We compared packing attenuation, initial occlusion, and recanalization rates between Cerecyte and bare platinum coils in the treatment of ruptured and unruptured cerebral aneurysms. MATERIALS AND METHODS We compared 63 patients (67 aneurysms) treated with Cerecyte coils with 65 patients (70 aneurysms) treated by using bare platinum coils. Results were classified by the Raymond score. Clinical outcomes were assessed by using a modified Rankin Scale. Angiographic and clinical follow-ups were performed routinely at 6 and 12 months after the intervention. RESULTS In the Cerecyte group, complete occlusion of the aneurysm (grade 1) was accomplished in 49% (33/67), a small residual neck (grade 2) was seen in 21% (14/67), and dome filling (grade 3) was seen in 30% (20/67). In the platinum group, 41% (29/70) were grade 1, 39% (27/70) were grade 2, and 20% (14/70) were grade 3 immediately postembolization. Mean packing attenuation was 43 +/- 28% in the Cerecyte group and 40 +/- 23% in the bare platinum group (P = .68). Twelve-month follow-up data were available for 54% (36/67) of the Cerecyte population and 43% (30/70) of the bare platinum population. There were 5 cases of neck recanalization (11%) in the Cerecyte group and 11 cases (23%) in the bare platinum group (P = .17). No rebleeds were noted in the follow-up period. CONCLUSIONS Cerecyte coils have a satisfactory safety profile. We were able to achieve high packing attenuations and initial occlusion rates similar to those obtainable with platinum coils.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cerecyte coils in the treatment of intracranial aneurysms: a preliminary clinical study.

BACKGROUND AND PURPOSE Recanalization remains a major drawback in the endovascular treatment of intracranial aneurysms. Here, we report on our preliminary clinical experience with a new bioactive coil. PATIENTS AND METHODS In a prospective study, 25 aneurysms were treated in 24 patients. Endovascular occlusion of the aneurysm was performed by using a novel polyglycol acid-loaded coil (Cerecyt...

متن کامل

Cerecyte coil trial: procedural safety and clinical outcomes in patients with ruptured and unruptured intracranial aneurysms.

BACKGROUND AND PURPOSE This study arose from a need to systematically evaluate the clinical and angiographic outcomes of intracranial aneurysms treated with modified coils. We report the procedural safety and clinical outcomes in a prospective randomized controlled trial of endovascular coiling for ruptured and unruptured intracranial aneurysms, comparing polymer-loaded Cerecyte coils with bare...

متن کامل

Cerecyte coil trial: angiographic outcomes of a prospective randomized trial comparing endovascular coiling of cerebral aneurysms with either cerecyte or bare platinum coils.

BACKGROUND AND PURPOSE We report the primary outcome of the Cerecyte Coil Trial, a randomized trial to determine whether polymer-loaded Cerecyte coils compared with Micrus bare platinum coils improved the proportion of patients with angiographic occlusion of the aneurysm at 6 months when assessed by a core laboratory. The secondary objectives were to compare the clinical outcomes and retreatmen...

متن کامل

Results of 101 aneurysms treated with polyglycolic/polylactic acid microfilament nexus coils compared with historical controls treated with standard coils.

BACKGROUND AND PURPOSE Polyglycolic/polylactic acid (PGLA) addition to bare platinum coils is intended to reduce the reopening rate of coiled intracranial aneurysms. Nexus coils are standard complex platinum coils with interwoven PGLA microfilament threads. We present the clinical results of 101 intracranial aneurysms treated with Nexus coils. MATERIALS AND METHODS Results of coiling of 101 a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 30 8  شماره 

صفحات  -

تاریخ انتشار 2009