Temperature-guided radiofrequency catheter ablation with very large distal electrodes.
نویسندگان
چکیده
BACKGROUND Previous studies have shown that the size of lesions produced by radiofrequency catheter ablation correlates with the temperature and surface area of the electrode-tissue interface. The purpose of the present study was to compare the effects of ablation using very large distal electrodes (8F, 8 and 12 mm long) with those made by a conventional radiofrequency ablation catheter (distal electrode 8F, 4 mm long). METHODS AND RESULTS Each catheter had a thermistor in the tip of the distal electrode. Radiofrequency energy (500 kHz) was supplied by a generator that continuously monitored temperature and produced up to 100 W. In 10 dogs, each of the three ablation catheters were introduced percutaneously and positioned under fluoroscopic guidance at disparate left ventricular endocardial sites. Radiofrequency power output was titrated to achieve a temperature of 80 degrees C for 60 seconds at each ablation site. The power required to produce a steady-state temperature of 80 degrees C was directly proportional to electrode size (15 +/- 7, 46 +/- 15, and 62 +/- 32 W using the 4-mm-, 8-mm-, and 12-mm-long electrodes, respectively). Lesions produced by the 8-mm electrode were nearly twice as deep (11 +/- 2.4 versus 6 +/- 1.2 mm, P < .001) and four times as large (905 +/- 410 versus 210 +/- 100 mm3, P < .001) as those made with a conventional 4-mm electrode. Lesions produced by the 12-mm electrode were intermediate in size (depth, 8 +/- 1.2 mm; volume, 465 +/- 225 mm3) and sometimes were associated with charring and crater formation. Ablation with the larger electrodes caused a drop in arterial pressure and more ventricular ectopy than ablation using a 4-mm distal electrode. CONCLUSIONS Thermistor-equipped elongated ablation electrodes coupled to high-power outputs can reproducibly produce lesions approximately 1 cm in diameter. This system may prove useful for ablation of ventricular tachycardias in patients with coronary artery disease.
منابع مشابه
Catheter Ablation of Atrioventricular Junction Using Radiofrequency Current in 17 Patients
Background. Two catheter electrode systems were compared for delivering radiofrequency current for ablation of the atrioventricular junction. Seventeen patients with drug-resistant supraventricular tachyarrhythmias were studied. Methods and Results. A 6F or 7F catheter with six or eight standard electrodes (1.25 mm wide, 2.5-mm spacing) was used in the first seven patients (group 1). A 7F quadr...
متن کاملMini Electrodes on Ablation Catheters: Valuable Addition or Redundant Information?—Insights from a Computational Study
Radiofrequency ablation has become a first-line approach for curative therapy of many cardiac arrhythmias. Various existing catheter designs provide high spatial resolution to identify the best spot for performing ablation and to assess lesion formation. However, creation of transmural and nonconducting ablation lesions requires usage of catheters with larger electrodes and improved thermal con...
متن کاملTemperature-controlled cooled-tip radiofrequency linear ablation of the atria guided by a realtime position management system.
Due to the difficulty in producing a transmural linear lesion and the possibility of complications such as thrombus formation leading to thromboembolism, the catheter-based maze procedure remains problematic. We tested, in pigs, the possibility of using a temperature-controlled cooled-tip radiofrequency (RF) ablation system together with a realtime position management (RPM) system to create a t...
متن کاملComputational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction
Radiofrequency catheter ablation (RFCA) is a routine treatment for cardiac arrhythmias. During RFCA, the electrode-tissue interface temperature should be kept below 80 °C to avoid thrombus formation. Open-irrigated electrodes facilitate power delivery while keeping low temperatures around the catheter. No computational model of an open-irrigated electrode in endocardial RFCA accounting for both...
متن کاملCatheter Ablation of Accessory Pathways
Background. Animal studies have suggested that the temperature of the electrode-tissue interface during radiofrequency catheter ablation accurately predicts lesion size. The purpose of the current study was to evaluate the utility of continuous temperature monitoring during radiofrequency catheter ablation in patients with Wolff-Parkinson-White syndrome. Methods and Results. Twenty patients wit...