Introduction Parallel transmission allows compensating for transmit field inhomogeneities in high field MRI by means of RF shimming, i.e. by the optimization of the amplitudes and phases of a multi-element RF array [1]. Moreover, the RF transmit power can be simultaneously reduced by regularization techniques. This allows for a reduction of the specific absorption rate (SAR) and hence reduces thermal stress for the patient. Unfortunately, the regularization approach leads to a conflict of objectives between image homogeneity and SAR reduction [2] and a compromise has to be found. So far, the RF transmission has been considered independently from the sampling process. It is well known, that the center of the receive k-space determines the image contrast and uniformity. This has been used for dynamic studies by employing the ‘keyhole’ concept [3]. For parallel transmission, this relation offers a new degree of freedom in the sequence design: Instead of using a fixed RF pulse for all phase-encoding steps, several different pulses can be applied [4]. This k-space adaptive approach to RF shimming avoids the trade-off between contrast and SAR. In the present work, this technique is discussed theoretically and proofof-principle is given based on phantom and in vivo images. Methods Imagine the 2D phase-encoding space of a 3D imaging sequence as shown in Fig. 1. In the central kspace, highly uniform, but relatively SAR-intensive RF pulses are to be used to achieve optimal contrast in the image domain. When sampling the outer k-space, the B1 homogeneity requirement can be relaxed to achieve a reduced average SAR. Since MR imaging is highly sensitive to phase inconsistencies, it is important to avoid deviations in the excitation phase when switching between different RF pulses. Therefore, a hybrid optimization method was developed: The central RF pulse is calculated using a magnitude least-squares (MLS) optimization that does not constrain the image phase [5]. The outer pulses are computed by a leastsquares (LS) optimization, using the Tx-phase obtained from the MLS step as the target phase. Experiments were performed on a 3T scanner (Philips Healthcare, Best, The Netherlands) equipped with an 8 TX-channel body coil [6] using a cylindrical water phantom (Ø40 cm) and in vivo on five average-sized volunteers in the abdominal region. B1-maps were acquired during breathhold for all Tx-channels using the actual flip angle (AFI) sequence with all-but-one channel transmitting to avoid regions of low flip angles [7]. RF-power (sum of the squared amplitude scaling factors) of the central pulses was slightly regularized to avoid exceedingly high drive voltages. Using the hybrid optimization method, additional pulses were calculated while step-wise tightening the RF power regularization. All pulses were normalized to give the same average flip-angle as the quadrature drive. Finally, the actual images were acquired using a 3D FFE Cartesian sequence for all calculated RF shim-sets. The combination of the acquired raw data in the k-space was performed off-line. The center of the k-space was filled with samples from the image with the MLS-optimized pulse, the outer k-space data were replaced with samples from the maximally regularized LS optimized raw data at different mixing ratios r. For the in vivo data, the whole-body SAR was calculated using an FDTD simulation. Results and Discussion Fig. 2 shows the RF-power plotted over the NRMSE (normalized root mean square error) of the effective B1 maps. The quadrature case is shown as a reference. Starting from a pure MLS-optimized excitation (r = 0%) and increasing r, the k-space adaptive design shows a strong reduction of RF-power, whereas homogeneity is almost completely maintained up to r = 80%. The corresponding images are shown in Figs. 3 and 4. When compared to the optimally shimmed images (a), the traditionally shimmed, low RF-power images (b) show a loss of contrast in the center. The k-space adaptive results (c) show an image quality comparable to the optimal results, while RF-power and SAR are almost halved. Conclusion A SAR reduction by k-space adaptive RF shimming is feasible without significant loss of image homogeneity. The conventional trade-off between the error and the RF-power can be overcome. In practice, using several RF pulses should allow for a smooth transition in the k-space when imaging in the steady-state. Furthermore, the proposed k-space adaptive shimming concept could be extended to all kinds of magnetization preparation RF pulses in more sophisticated imaging sequences such as TFE to further reduce the average SAR.
منابع مشابه
Slab selective, regularized RF shimming
Introduction RF shimming is a well-known technique, where amplitudes and phases of a transmit RF coil array are optimized to obtain a homogeneous B1 profile, i.e., to compensate wave propagation effects at main fields of 3T or above (see, e.g., [1]). Among other anatomies, legs are particularly prone to wave propagation effects due to their bi-cylindrical geometry, resulting in significant sign...
متن کاملMR Physics for Physicists / B1 Shimming and Parallel Transmission
METHODS: Currently, among other means, modifications of the applied RF pulses are investigated to cope with the mentioned B1 inhomogeneities. First, using a coil array for RF transmission, standard slice selective pulses can be transmitted with different weights in the different array elements (see, e.g., [2-4]). These weights, i.e., amplitude and phase of each element, are designed to yield a ...
متن کاملAverage SAR Constrained Local RF Shimming
Introduction Achieving given homogeneity and SAR constraints in RF shimming process is an important problem in high field MRI. There are different studies in the literature which dealt with the optimization of these parameters by using non-convex optimization techniques. In a previous work whole body SAR constrained RF shimming problem was solved in order to achieve whole slice transmit homogen...
متن کاملEigenmode Analysis of Transmit Coil Array for SAR-reduced B1 Mapping and RF shimming
Introduction The B1 transmit field inhomogeneity represents a serious problem in whole-body high field MRI (≥3T). B1 shimming based on measured B1 maps is a promising approach to cope with this problem and represents the primary application for parallel transmission at this point in time [1,2]. However, B1 mapping is still an error-prone and time consuming process [3], potentially resulting in ...
متن کاملElectrodynamic analysis of SAR and transmit homogeneity for RF shimming on a dielectric cylinder
Introduction In RF shimming [1,2], a single RF current waveform is modulated in amplitude and phase at each coil of a transmit array, in order to control the distribution of electromagnetic (EM) fields and compensate for B1 inhomogeneities. An algorithm to minimize the specific absorption rate (SAR) for a given target profile using RF shimming has been recently proposed and evaluated in simulat...