Prosthetic heart valves and annuloplasty rings: assessment of magnetic field interactions, heating, and artifacts at 1.5 Tesla.

نویسنده

  • F G Shellock
چکیده

The purpose of this study was to determine the magnetic resonance (MR) safety aspects and artifacts for three different heart valve prostheses and two different annuloplasty rings that have not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection angle and torque), heating (gel-filled phantom and fluoroptic thermometry; 15 min of MR imaging at a whole body-averaged specific absorption rate of 1.2 W/kg), and artifacts (using T1-weighted, spin echo, and gradient echo pulse sequences). One heart valve prosthesis and one annuloplasty ring showed no magnetic field interactions. Two heart valve prostheses and one annuloplasty ring displayed relatively minor magnetic field interactions (i.e., deflection angle < or = 6 degrees, torque, +1). Heating was < or = 0.7 degrees C for the five different implants. Artifacts varied depending on the amount and type of metal used to make the implants. For the three heart valve prostheses and two annuloplasty rings, the lack of substantial magnetic field interactions and relatively minor hearing indicated that MR procedures may be conducted safetly in patients with these implants using MR systems operating with static magneticfields of 1.5 T or less. Notably, these findings essentially apply to 54 different heart valve prostheses and 37 different annuloplasty rings (i.e., based on the various models and sizes available for these implants).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Prosthetic heart valves: evaluation of magnetic field interactions, heating, and artifacts at 1.5 T.

The purpose of this study was to use ex vivo testing techniques to determine the magnetic resonance imaging (MRI) safety aspects for 32 different heart valve prostheses that had not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection angle and tor...

متن کامل

Assessment of MRI Issues at 3 Tesla for a New Metallic Tissue Marker

Purpose. To assess the MRI issues at 3 Tesla for a metallic tissue marker used to localize removal areas of tissue abnormalities. Materials and Methods. A newly designed, metallic tissue marker (Achieve Marker, CareFusion, Vernon Hills, IL) used to mark biopsy sites, particularly in breasts, was assessed for MRI issues which included standardized tests to determine magnetic field interactions (...

متن کامل

Magnetically programmable shunt valve: MRI at 3-Tesla.

A magnetically programmable cerebrospinal fluid (CSF) shunt valve (Codman Hakim Programmable Valve, Codman, a Johnson & Johnson Company, Raynham, MA) was assessed for magnetic field interactions, heating, artifacts and functional changes at 3-Tesla. The programmable valve showed minor magnetic field interactions and heating (+0.4 degrees C). Artifacts were relatively large in relation to the si...

متن کامل

Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla

PURPOSE To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). METHODS The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso ...

متن کامل

Detachable coil for cerebral aneurysms: in vitro evaluation of magnetic field interactions, heating, and artifacts at 3T.

A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this impl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance

دوره 3 4  شماره 

صفحات  -

تاریخ انتشار 2001