Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography

نویسندگان

  • Onur Polat
  • Zeki Baysal
  • Serkan Özcan
  • Sibel İnan
  • Ümit Übeyt İnan
چکیده

OBJECTIVES To assess the agreement of anterior segment parameter measurements derived from Aladdin optical biometer using optical low coherence interferometer and Sirius corneal topography using combined Scheimpflug-Placido disk. MATERIALS AND METHODS Data obtained using the Aladdin and Sirius systems from 110 eyes of 59 subjects who had no health problems other than refractive errors were retrospectively evaluated. Anterior chamber depth (ACD), flat (K1) and steep (K2) keratometry readings, and white-to-white distance (WTW) measurements taken with both devices were noted. RESULTS The mean age of the patients was 47.31±18.57 years (range, 25 to 79 years). Mean ACD was 3.35±0.4 mm using Aladdin and 3.42±0.44 mm using Sirius. Mean difference in ACD was 0.075 mm greater with Sirius than Aladdin (p<0.001). K1 measurement obtained by Aladdin was an average of 0.409 D higher (p<0.001). No statistically significant differences were detected between the two devices in respect to K2 and WTW measurements (p=0.18, p=0.85 respectively). Pearson correlation analysis showed high correlation between the two devices for all measurements (r=0.985, 0.895, 0.961 and 0.766 for ACD, K1, K2 and WTW respectively; p<0.001). CONCLUSION Anterior segment parameters obtained by Aladdin optical biometer and Sirius anterior segment analysis system correlated well with each other and measurement differences between the devices were clinically negligible except for K1 values.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts

Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), ...

متن کامل

Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology

PURPOSE The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-...

متن کامل

Changes in the anterior segment after cycloplegia with a biometer using swept-source optical coherence tomography

The aim of this study was to investigate changes in the anterior segment of the eye after cycloplegia. A biometer combined with swept-source optical coherence tomography (SSOCT) was used for measurements. Patients with strabismus or amblyopia who underwent cycloplegia were included. The axial length, central corneal thickness, anterior chamber depth, and lens thickness were measured with the bi...

متن کامل

Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation

PURPOSE To investigate the accuracy of a new optical biometer for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery. METHODS Consecutive eyes of patients undergoing cataract surgery with the same IOL model were enrolled in a prospective cohort study. Axial length (AL) and corneal power were measured with an optical biometer based on optical low-coherence interferome...

متن کامل

Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug + Placido device

PURPOSE To analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters. METHODS In this observational study, 100 normal candidates underwent Scheimpflug analysis with Sirius 3D Rotating Scheimpflug Camera and Topography System (Costruzione Strumenti Oftalmici, Italy). Two scans in dark room conditions were performed by the same experi...