Mechanical failure of the Intramedullary Skeletal Kinetic Distractor in limb lengthening.

نویسندگان

  • R D Burghardt
  • J E Herzenberg
  • S C Specht
  • D Paley
چکیده

Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Femoral lengthening with the Intramedullary Skeletal Kinetic Distractor.

Femoral lengthening using the Intramedullary Skeletal Kinetic Distractor is a new technique. However, with intramedullary distraction the surgeon has less control over the lengthening process. Therefore, 33 femora lengthened with this device were assessed to evaluate the effect of operative variables under the surgeon's control on the course of lengthening. The desired lengthening was achieved ...

متن کامل

Limb-length discrepancy after total hip arthroplasty: novel treatment and proposed algorithm for care.

Limb-length discrepancy after total hip arthroplasty (THA) leads to patient dissatisfaction and can be a cause of orthopedic surgery malpractice cases. Nonsurgical and surgical techniques exist to correct limb-length discrepancies. Two limb-lengthening methods were used to correct greater than 2-cm limb-length discrepancies after THA: lengthening over a femoral nail with an external fixator and...

متن کامل

Intramedullary limb lengthening with the Intramedullary Skeletal Kinetic Distractor in the lower limb.

We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley's classification for evaluating proble...

متن کامل

The effect on mechanical axis deviation of femoral lengthening with an intramedullary telescopic nail.

Internal lengthening devices in the femur lengthen along the anatomical axis, potentially creating lateral shift of the mechanical axis. We aimed to determine whether femoral lengthening along the anatomical axis has an inadvertent effect on lower limb alignment. Isolated femoral lengthening using the Intramedullary Skeletal Kinetic Distractor was performed in 27 femora in 24 patients...

متن کامل

The intramedullary skeletal kinetic distractor (ISKD): first clinical results of a new intramedullary nail for lengthening of the femur and tibia.

In 1986, a programme was initiated by the senior author to develop a reliable, mechanically activated, intramedullary lengthening device with a non-invasive means of measuring the progress of lengthening without X-ray. We report results of design, biomechanical testing, in vivo animal testing and clinical implantation of the first 20 intramedullary skeletal kinetic distractors (ISKDs) in adult ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 93 5  شماره 

صفحات  -

تاریخ انتشار 2011