MR imaging of patients with intracranial aneurysm clips.

نویسندگان

  • E Kanal
  • F G Shellock
چکیده

Four patients with intracranial aneurysm clips made from a variety of alloys were studied without incidence by MR imaging at field strengths ranging from 0.35 to 0.6 T. Knowledge of the type of alloy used in the manufacturing of an aneurysm clip is important in determining whether the clip will or will not deflect in a magnetic field. Ferromagnetic clips show deflection and torque in a magnetic field and have the potential to dislodge from the aneurysm. Nonferromagnetic or weakly ferromagnetic aneurysm clips such as the Sugita (Elgiloy), Yasargil (316 LVM stainless steel), Heifetz (Elgiloy), Yasargil (Phynox), and Vari-Angle McFadden (MP35N) do not deflect or deflect weakly in the magnetic field and therefore would not be expected to dislodge during MR. The option of imaging many patients with intracranial aneurysm clips with MR extends the usefulness of the technique to a previously excluded population.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of MR imaging and CT in patients with intracranial aneurysm clips.

CT and MR imaging of the brain were performed without complications in 16 patients who had undergone surgery for a ruptured aneurysm with the use of nonferromagnetic Yasargil 316 or Sugita Elgiloy clips. MR was performed on a 0.3-T Fonar beta-3000M imaging system. The artifacts caused by the clips were smaller on MR than on CT, and, therefore, anatomic structures such as the brainstem and tempo...

متن کامل

Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

BACKGROUND AND PURPOSE The use of 3.0-T MR systems is increasing worldwide. We evaluated magnetic field interactions and translational attraction for 32 aneurysm clips in association with exposure to "long-bore" and "short-bore" 3.0-T MR imaging systems. METHODS Thirty-two different aneurysm clips were evaluated in this investigation. Each aneurysm clip was qualitatively evaluated for magneti...

متن کامل

Aneurysm clips: evaluation of magnetic field interactions with an 8.0 T MR system.

This study was conducted to evaluate magnetic field interactions for aneurysm clips exposed to an 8.0 T magnetic resonance (MR) system. Twenty-six different aneurysm clips were tested for magnetic field translational attraction (deflection angle test) and torque (qualitative assessment method) using previously described techniques. Six of the specific aneurysm clips (i.e. type, model, blade len...

متن کامل

Comparison of MR Imaging and CT in Patients with lntracranial Aneurysm

CT and MR imaging of the brain were performed without complications in 16 patients who had undergone surgery for a ruptured aneurysm with the use of nonferromagnetic Yasargil 316 or Sugita Elgiloy clips. MR was performed on a 0.3-T Fonar ,B-3000M imaging system. The artifacts caused by the clips were smaller on MR than on CT, and, therefore, anatomic structures such as the brainstem and tempora...

متن کامل

In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

BACKGROUND AND PURPOSE Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imaging-related heating, and artifacts. Therefore, we evaluated these risks for aneurysm clips. MATERIALS AND METHODS Three different MP35N aneurysm clips (Codman Slim-Line Aneurysm Clip, straight, blade length 25-mm; Codman Slim-Line Aneurysm Clip Graft...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 9 5  شماره 

صفحات  -

تاریخ انتشار 1988