Moving gantry method for electron beam dose profile measurement at extended source‐to‐surface distances

نویسندگان

  • Gábor Fekete
  • Emese Fodor
  • Csilla Pesznyák
چکیده

A novel method has been put forward for very large electron beam profile measurement. With this method, absorbed dose profiles can be measured at any depth in a solid phantom for total skin electron therapy. Electron beam dose profiles were collected with two different methods. Profile measurements were performed at 0.2 and 1.2 cm depths with a parallel plate and a thimble chamber, respectively. 108cm × 108 cm and 45 cm × 45 cm projected size electron beams were scanned by vertically moving phantom and detector at 300 cm source-to-surface distance with 90° and 270° gantry angles. The profiles collected this way were used as reference. Afterwards, the phantom was fixed on the central axis and the gantry was rotated with certain angular steps. After applying correction for the different source-to-detector distances and incidence of angle, the profiles measured in the two different setups were compared. Correction formalism has been developed. The agreement between the cross profiles taken at the depth of maximum dose with the 'classical' scanning and with the new moving gantry method was better than 0.5 % in the measuring range from zero to 71.9 cm. Inverse square and attenuation corrections had to be applied. The profiles measured with the parallel plate chamber agree better than 1%, except for the penumbra region, where the maximum difference is 1.5%. With the moving gantry method, very large electron field profiles can be measured at any depth in a solid phantom with high accuracy and reproducibility and with much less time per step. No special instrumentation is needed. The method can be used for commissioning of very large electron beams for computer-assisted treatment planning, for designing beam modifiers to improve dose uniformity, and for verification of computed dose profiles.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Generating a Homogeneous Dose Distribution in the Junction Region between Two Adjacent Fields in Electron Beam Therapy

Introduction: Treatment with megavoltage electron beams is ideal for irradiating superficial tumors because of their limited range in tissues. However, for electron treatment of extended areas, such as the chest wall, two or more adjacent fields can be used. Abutment of these fields may lead to significant dose in homogeneities in the junction region. The aim of this study is to offer a new met...

متن کامل

Impact of Silicone gel breast prosthesis on photon beam therapy of recurrence carcinoma post mastectomy

AbstractIntroduction Silicone gel breast implants are used for breast reconstruction post mastectomy. In the event of cancer recurrence, the radiation oncologists are forced to irradiate through the prosthesis device. Due to prosthesis higher atomic number dose perturbations occur during treatment. This study determined the influence of silicone gel thickness on the photon beam distribution. Ma...

متن کامل

Benchmarking of Monte Carlo model of 6 Mv photon beam produced by Siemens Oncor® linear accelerator: determination of initial electron beam parameters in comparison with measurement

Introduction: The aim of this study was to investigate the initial electron beam parameters for Monte Carlo model of 6MV photon beam produced by Siemens Oncor® linear accelerator.   Materials and Methods: In this study, the EGSnrc Monte Carlo user codes BEAMnrc and DOSXYZnrc were used. The beamnrc code were used for modelling of a 6 MV photon beam produced by...

متن کامل

Electron beam quality control using an amorphous silicon EPID.

An amorphous silicon EPID has been investigated to determine whether it is capable of quality control constancy measurements for linear accelerator electron beams. The EPID grayscale response was found to be extremely linear with dose over a wide dose range and, more specifically, for exposures of 95-100 MU. Small discrepancies of up to 0.8% in linearity were found at 6 MeV (8-15 MeV showed bet...

متن کامل

A Monte Carlo evaluation of beam characteristics for total body irradiation at extended treatment distances

The aim is to study beam characteristics at large distances when focusing on the electron component. In particular, to investigate the utility of spoilers with various thicknesses as an electron source, as well as the effect of different spoiler-to-surface distances (STSD) on the beam characteristics and, consequently, on the dose in the superficial region. A MC model of a 15 MV Varian accelera...