Insufficiency in a new temporal-artery thermometer for adult and pediatric patients.

نویسندگان

  • Mohammad-Irfan Suleman
  • Anthony G Doufas
  • Ozan Akça
  • Michel Ducharme
  • Daniel I Sessler
چکیده

UNLABELLED SensorTouch is a new noninvasive temperature monitor and consists of an infrared scanner that detects the highest temperature on the skin of the forehead, presumably over the temporal artery. The device estimates core temperature (T(core)). We tested the hypothesis that the SensorTouch is sufficiently precise and accurate for routine clinical use. We studied adults (n = 15) and children (n = 16) who developed mild fever, a core temperature of at least 37.8 degrees C, after cardiopulmonary bypass. Temperature was recorded at 15-min intervals throughout recovery with the SensorTouch thermometer and from the pulmonary artery (adults) or bladder (children). Pulmonary artery (T(core)) and SensorTouch (T(st)) temperatures correlated poorly in adults: T(core) = 0.7. T(st) + 13, r(2) = 0.3. Infrared and pulmonary artery temperatures differed by 1.3 +/- 0.6 degrees C; 89% of the adult temperatures thus differed by more than 0.5 degrees C. Bladder and infrared temperatures correlated somewhat better in pediatric patients: T(core) = 0.9. T(st) + 12, r(2) = 0.6. Infrared and bladder temperatures in children differed by only 0.3 degrees C, but the SD of the difference was 0.5 degrees C. Thus, 31% of the values in the infants and children differed by more than 0.5 degrees C. IMPLICATIONS We evaluated a noninvasive infrared forehead thermometer (SensorTouch) in adult and pediatric cardiac patients. Accuracy was poor in the adults and suboptimal in infants and children.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison sensitivity and specificity temporal artery temperature measurement with axillary and nasopharyngeal methods in adult

      Aims: temporal artery thermometers (TAT) of new tools for measuring body temperature, few studies have been conducted to evaluate the accuracy and reliability of thermometers and is still uncertain accuracy and reliability this thermometers.This study to evaluate the accuracy of the two methods of measuring body temperature by the thermometer axillary and temporal artery compared with nas...

متن کامل

Accuracy of a noninvasive temporal artery thermometer for use in infants.

OBJECTIVES To assess the accuracy of a new noninvasive temporal artery (TA) thermometer in infants; to compare the accuracy of the TA thermometer with that of a tympanic thermometer, using rectal thermometry as the criterion standard; and to compare the tolerability of the TA thermometer with that of the tympanic and rectal thermometers. DESIGN Prospective evaluation of the accuracy of TA and...

متن کامل

Main indications and long-term outcomes of reoperation after initial repair of tetralogy of Fallot

Background and Objective: The aim of this study was to analyze our indications, surgical procedures, and clinical outcomes of patients undergoing reoperation after surgical correction of tetralogy of Fallot (TOF). Methods: Thirty seven consecutive patients who underwent reoperation late after intra-cardiac repair of TOF within a period of 10 years were assessed. Results: The most co...

متن کامل

Temporal Artery versus Bladder Thermometry during Adult Medical-Surgical Intensive Care Monitoring: An Observational Study

BACKGROUND We sought to evaluate agreement between a new and widely implemented method of temperature measurement in critical care, temporal artery thermometry and an established method of core temperature measurement, bladder thermometry as performed in clinical practice. METHODS Temperatures were simultaneously recorded hourly (n = 736 observations) using both devices as part of routine cli...

متن کامل

Temporal artery thermometry in children younger than 5 years: a comparison with rectal thermometry.

BACKGROUND Temporal artery (TA) thermometry has come as one of the new methods for temperature measurement, especially in children in whom accurate temperature monitoring can save lives. The device which is convenient and simple to use is yet to gain popularity in several parts of the world, as there are conflicting reports of its accuracy. This study compares the accuracy of the TA thermometry...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Anesthesia and analgesia

دوره 95 1  شماره 

صفحات  -