Towards dynamic shimming in a 31cm bore 9.4T system: analysis of shim-shim inductive interactions

نویسندگان

  • D. W. Haw
  • B. A. Chronik
چکیده

Introduction: Preclinical MRI and spectroscopy applications at very high field demand improved shimming capabilities. Our goal is to improve shimming techniques in a 31cm bore 9.4T MR scanner with the development of an optimized high strength dynamic shim system. There are two types of challenges surrounding dynamic shimming: one is shim-shim interactions, and the other is interactions between the shim coils and the rest of the system. Attempts have been made to alleviate the problem of shim-shim interactions with limited success [1]. In this abstract we report on the severity of shim-shim interactions for all axes up to 2 order in tesseral and 3 order in zonal. We also investigate a method to reduce the interactions between shims that exhibit significant coupling. Methods: Shim axes of various orders and predetermined fixed lengths were designed using a Fourier series method [2]. This method expresses current density over a specified interval as a truncated Fourier series. Outside the interval the current density is set to zero. Magnetic field and resistance are expressed in terms of current density, and a functional that includes these coil parameters is minimized. This algorithm was first used to design a realistic shim set containing: Z, X, Y, Z, XY, XZ, YZ, X-Y, Z, and Z. All shim coils were individually designed using a minimum power technique. A mutual inductance term was then added to the functional and used to design decoupled shims. The inner most shim coil, Z, has a radius of 10cm. The radius of each coil, preceding radially outward, is 1cm larger than the previous shim coil. All coils had a fixed aspect ratio (diameter/length) of 2.5, a self-inductance of 200μH and were designed to have the same field uniformity over a 10cm DSV. The mutual inductances between all shim axes were calculated using a modified Neumann formula based on realistic discrete wire patterns. For mutual inductance values deemed to be significant, the higher order shim coil of the interacting pair was re-designed using the minimum mutual inductance technique. Results and Discussion: It was determined that the majority of shim axes do not posses any significant interactions between them. In fact the only significant interactions seen were between the Z and Z shims, and the Z and Z shims. The mutual inductance values for these two sets of interacting shims, along with the re-designed decoupled Z and Z shims are given in table 1. All other mutual inductance values were a factor of 10 less than the coupling zonal shims, and thus were deemed to be insignificant. It was also found that the new decoupled shims did not interact in any significant way with the other original shim axes. Figures 1 and 2 show the wire patterns for the original minimum power Z-Z and Z-Z interacting pairs and the corresponding re-designed minimum mutual inductance Z and Z shims. It can be seen that the re-designed Z and Z shims couple to the Z and Z coils by factors of 50 and 25 less than the original designs. The effect of this reduced coupling is significant. For example, given the original Z-Z set, if the current through the Z shim is changed by driving with a 400V source, the induced voltage on the Z coil is 146.2V. However, for the decoupled Z shim the induced voltage would only be 2.92V, well within the ability of most power supplies to compensate. It has been shown that mutual inductive interactions between realistic shim axes are dominated by only a few coil combinations. Furthermore, it is possible to redesign these axes to reduce this interaction to manageable levels. References: [1] de Graaf R A et al. 2003 Magn. Reson. Med. 49 409-416 [2] Carlson J W et al. 1992 Magn. Reson. Med. 26 191-206 Figure 1. Minimum power Z shim coil (top). Minimum power Z shim coil (middle). Minimum mutual inductance Z shim coil (bottom).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Shielding requirements for a complete 3 order shim set for a 31cm bore 9.4T system

Introduction: MRI and in-vivo spectroscopy in preclinical applications at very high field (9.4T and above) require improved shimming capabilities. Our goal is to develop and optimize high-strength, dynamic shim systems for our 31cm bore 9.4T MR system. It is clear that dynamic shim systems will need to be actively shielded [1], especially higher order shims, yet no attempts have been made to ad...

متن کامل

System Identification and Signal Processing for PID Control of B0 Shim Systems in Ultra-High Field Magnetic Resonance Applications

Magnetic resonance (MR) scanners are important tools in medical diagnostics and in many areas of neuroscience. MR technology is moving towards ultra-high field (UHF) 7T and 9.4T scanners which provide more signal intensity. However they also suffer from inhomogeneity of the static (B0) magnetic field which can lead to artifacts and uninterpretable data. B0 shimming is a technique used to reduce...

متن کامل

A Simple Approach of Combined Passive and Active Shimming for In Vivo MR Spectroscopy at High Magnetic Field

High magnetic fields increase the sensitivity and spectral dispersion in MR spectroscopy. However, spectral peaks are broadened in vivo at higher field strength due to stronger susceptibility-induced effects. Strategies to minimize the spectral linewidth by optimal shimming are therefore of critical importance. In our current 7T MR system, the shim requirements exceeded the dynamic range of the...

متن کامل

Dynamic Bo Shimming at 7 Tesla

Introduction Dynamic Shimming is a technique for obtaining optimal Bo field homogeneity over a volume by updating the shim coil and gradient currents for every slice in a multislice acquisition in real time [1, 2]. Dynamic shimming can produce better field homogeneity within each slice than global volume shimming methods. It can therefore reduce signal losses and geometric distortions and can b...

متن کامل

The effect of high-permittivity pads on specific absorption rate in radiofrequency-shimmed dual-transmit cardiovascular magnetic resonance at 3T

BACKGROUND Dual-channel transmit technology improves the image quality in cardiovascular magnetic resonance (CMR) at 3 T by reducing the degree of radiofrequency (RF) shading over the heart by using RF shimming. Further improvements in image quality have been shown on a dual-transmit system using high permittivity pads. The aim of this study is to investigate the transmit field (B 1 (+)) homoge...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008