Size estimation and magnification error in radiographic imaging: implications for classification of arteriovenous malformations.
نویسندگان
چکیده
PURPOSE To assess magnification error in digital subtraction angiography as it pertains to arteriovenous malformation (AVM) size. METHODS A rectangular grid phantom with equally spaced markers mounted in a stereotactic frame was imaged with digital angiographic equipment. The location and orientation of the grid was altered relative to the central plane of the phantom. Both linear and area measurements were made according to the perceived location of phantom markers using a standard catheter calibration technique and compared with stereotactically derived estimates. Finally, a single case example of an angiographically imaged rolandic AVM was used to compare linear dimensions obtained with both described techniques. RESULTS The determination of location and size with standard angiographic imaging is subject to error because of the divergent geometry of the incident x-ray beam. The resulting nonconstant geometric magnification causes errors in linear measurements of 10% to 13% at depths of 7 cm from the calibration plane. Errors in area measurements at the same position increase by 20% to 25%. Measurements of maximum diameter or cross-sectional area may have an additional error when nonspherical objects are inclined to the viewing direction (40% at 45 degrees inclination). These errors are reduced to less than 1 mm using the stereotactic technique. Some commercial angiographic systems have internal software to enable a spatial calibration based on known distances in the image or on the diameter of a catheter. The catheter technique was accurate in the calibration direction (perpendicular to the catheter axis) but had a 12% error in the direction parallel to the catheter because of a nonunity aspect ratio in the video system. Measurement of the dimensions of a rolandic AVM using the catheter calibration technique had an error that ranged from -3% to +26% (standard error, 20%) with respect to the stereotactic technique. CONCLUSIONS Numerous nonstereotactic referential systems for determining linear distances are inherently erroneous by varying degrees compared with the stereotactic technique. Area and volume determinations naturally increase this error further. To the extent that no standardized method for determining linear distances exists, significant variations in estimation of AVM size result. Classification schemes for AVMs have been hampered by this technical error.
منابع مشابه
دقت رادیوگرافی پانورامیک و توموگرافی خطی در تعیین محل کانال مندیبولار
Background and Aim: Accurate bone measurements are essential to determine the optimal size and length of dental implants. The magnification factor of radiographic images may vary with the imaging technique used. The purpose of this study was to compare the accuracy of linear tomography and panoramic radiography in vertical measurements, as well as the accuracy of linear tomography in mandibular...
متن کاملبررسی دقت توموگرافی خطی در تعیین محل حفره بینی و سینوس ماگزیلاری1
Background and Aim: Accurate measurement of bone height and width is essential prior to dental implant placement. The method of surgery as well as, the type and size of implants are determined according to dimensions of the residual bone. The purpose of this study was to evaluate the accuracy of linear tomography in localization of the floor of nasal fossa and maxillary sinus, and to determine ...
متن کاملThe Synergistic Effect of Bleomycin, Triamcinolone and Epinephrine in Treatment of Hemangioma and Arteriovenous Malformations
BACKGROUND Hemangioma is the most common tumor of neonatal period and it is almost always appeared by the end of the first week of life and can be found mostly in head and neck area. This study evaluated the synergistic effect of bleomycin, triamcinolone and epinephrine in treatment of hemangioma and arte-riovenous malformations. METHODS In this multicenter study, a combination of bleomycin,...
متن کاملThe “Little AVM”: A New Entity in High-flow versus Low-flow Vascular Malformations
BACKGROUND Arteriovenous malformations (AVMs) are high-flow lesions with abnormal connections between arteries and veins without an intervening capillary bed. Infrequently, the radiographic diagnosis of a vascular lesion will not support the clinical diagnosis of an AVM. These "discrepant" lesions are not adequately captured within the current classification system and represent a treatment dil...
متن کاملHysterosalpingography in the assessment of congenital cervical Anomalies: Pictorial Review
Cervical abnormalities may be congenital or acquired. Congenital anomalies of the cervix are rarely isolated, and more commonly accompany other uterine anomalies. Various imaging tools have been used in the assessment of Mullerian duct anomalies (MDAs). Currently, magnetic resonance imaging (MRI) is the modality of choice for definitive diagnosis and classification of the MDAs. Hysterosalpingog...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- AJNR. American journal of neuroradiology
دوره 16 3 شماره
صفحات -
تاریخ انتشار 1995