A hazard analysis for a generic insulin infusion pump.
نویسندگان
چکیده
BACKGROUND Researchers at the Food and Drug Administration (FDA)/Center for Device and Radiological Health/Office of Science and Engineering Laboratories have been exploring the concept of model-based engineering as a means for improving the quality of medical device software. Insulin pumps were chosen as a research subject because their design provides the desired degree of research complexity and these types of devices present an ongoing regulatory challenge. METHODS Insulin pump hazards and their contributing factors are considered in the context of a highly abstract generic insulin infusion pump (GIIP) model. Hazards were identified by consulting with manufacturers, pump users, and clinicians; by reviewing national and international standards and adverse event reports collected by the FDA; and from workshops sponsored by Diabetes Technology Society. This information has been consolidated in tabular form to facilitate further community analysis and discussion. RESULTS A generic insulin infusion pump model architecture has been established. A fairly comprehensive hazard analysis document, corresponding to the GIIP model, is presented in this article. CONCLUSIONS We believe that this work represents the genesis of an insulin pump safety reference standard upon which future insulin pump designs can be based to help ensure a basic level of safety. More interaction with the diabetes community is needed to assure the quality of this safety modeling process.
منابع مشابه
Qualitative Application of Operation and Support Hazard Analysis for Syringe Infusion Pumps
In this paper, we provide a qualitative analysis with an experimental viewpoint for the Operation and Support Hazard Analysis O&SHA process applied to syringe infusion pumps as a cheap and accurate way of contributing with the mitigation of the safety problem in these devices. The proposed model takes as quality criteria the compliance within safety requirements defined by the Foods and Drugs A...
متن کاملA Generic User Interface Architecture for Analyzing Use Hazards in Infusion Pump Software
This paper presents a generic infusion pump user interface (GIP-UI) architecture that intends to capture the common characteristics and functionalities of interactive software incorporated in broad classes of infusion pumps. It is designed to facilitate the identification of use hazards and their causes in infusion pump designs. This architecture constitutes our first effort at establishing a m...
متن کامل5th Workshop on Medical Cyber-Physical Systems
This paper presents a generic infusion pump user interface (GIP-UI) architecture that intends to capture the common characteristics and functionalities of interactive software incorporated in broad classes of infusion pumps. It is designed to facilitate the identification of use hazards and their causes in infusion pump designs. This architecture constitutes our first effort at establishing a m...
متن کاملInsulin pump therapy, multiple daily injections, and cardiovascular mortality in 18 168 people with type 1 diabetes: observational study
OBJECTIVE To investigate the long term effects of continuous subcutaneous insulin infusion (insulin pump therapy) on cardiovascular diseases and mortality in people with type 1 diabetes. DESIGN Observational study. SETTING Swedish National Diabetes Register, Sweden 2005-12. PARTICIPANTS 18,168 people with type 1 diabetes, 2441 using insulin pump therapy and 15,727 using multiple daily ins...
متن کاملEffect of 12-Week Body Pump with Consumption of Cinnamon and Honey on Resistin, Visfatin and Insulin Resistance in Overweight Children
Background and Objectives: Prevention of obesity and overweight and the associated risk factors such as insulin resistance from an early age, is of particular importance. The aim of this study was to investigate the effect of 12-week Body Pump and combination of Cinnamon and Honey on resistin, visfatin and insulin resistance in overweight children. Materials and Methods: The present applied an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of diabetes science and technology
دوره 4 2 شماره
صفحات -
تاریخ انتشار 2010