Comparative evaluation of Löwenstein-Jensen proportion method, BacT/ALERT 3D system, and enzymatic pyrazinamidase assay for pyrazinamide susceptibility testing of Mycobacterium tuberculosis.
نویسندگان
چکیده
Pyrazinamide (PZA) is an important first-line antituberculosis drug because of its sterilizing activity against semidormant tubercle bacilli. In spite of its very high in vivo activity, its in vitro activity is not apparent unless an acidic environment is available, which makes PZA susceptibility testing difficult by conventional methods. The present study was, therefore, planned to assess the performance of the colorimetric BacT/ALERT 3D system and compare the results with those from conventional tests, i.e., the Löwenstein-Jensen (LJ) proportion method (pH 4.85) and Wayne's pyrazinamidase (PZase) assay, using 107 clinical isolates. The concordance among all of these tests was 89.71% after the first round of testing and reached 92.52% after resolution of the discordant results by retesting. Prolonged incubation of the PZase tube for up to 10 days was found to increase the specificity of the PZase test. The concordances between LJ proportion and BacT/ALERT 3D, LJ proportion and the PZase assay, and BacT/ALERT 3D and the PZase assay were found to be 99.06%, 93.46%, and 92.52%, respectively. Using the LJ results as the gold standard, the sensitivities of BacT/ALERT 3D and the PZase assay were 100 and 82.85%, respectively, while the specificity was 98.61% for both of the tests. The difference between the sensitivities of BacT/ALERT 3D and the PZase assay was significant (P = 0.025). The mean turnaround times for the detection of resistant and susceptible results by BacT/ALERT 3D were 8.04 and 11.32 days, respectively. While the major limitations associated with the PZase assay and the LJ proportion method are lower sensitivity in previously treated patients and a longer time requirement, respectively, the BacT/ALERT 3D system was found to be rapid, highly sensitive, and specific.
منابع مشابه
Standardization of first and second-line antitubercular susceptibility testing using BacT Alert 3D system: a report from a tertiary care centre in India.
Resurgence of multidrug resistant tuberculosis has lead to demand for rapid susceptibility testing. Conventional methods take > 3 weeks and are tedious. Automated methods have superseded them for first line drug susceptibility testing. An attempt was made to standardize first and second line susceptibility testing using the BacT Alert 3D system (Biomerieux). And compare results with Lowenstein ...
متن کاملPerformance of the BacT Alert 3D System Versus Solid Media for Recovery and Drug Susceptibility Testing of Mycobacterium tuberculosis in a Tertiary Hospital in Korea
BACKGROUND Tuberculosis (TB) is a major health problem, and accurate and rapid diagnosis of multidrug-resistant (MDR) and extended drug-resistant (XDR) TB is important for appropriate treatment. In this study, performances of solid and liquid culture methods were compared with respect to MDR- and XDR-TB isolate recovery and drug susceptibility testing. METHODS Sputum specimens from 304 patien...
متن کاملEvaluation of the BacT/ALERT PZA kit in comparison with the BACTEC 460TB PZA for testing Mycobacterium tuberculosis susceptibility to pyrazinamide.
OBJECTIVES To compare the performance of the BacT/ALERT PZA kit (BioMerieux, Marcy l'Etoile, France) with the radiometric BACTEC 460TB PZA test (Becton-Dickinson method) for testing Mycobacterium tuberculosis susceptibility to pyrazinamide. METHODS A total of 50 M. tuberculosis strains were tested. Thirty of these strains had been previously considered pyrazinamide-susceptible and 20 pyrazina...
متن کاملEvaluation of BacT/Alert 3D liquid culture system for recovery of mycobacteria from clinical specimens using sodium dodecyl (lauryl) sulfate-NaOH decontamination.
A total of 52 mycobacterial isolates were recovered from 1,197 clinical specimens decontaminated by a sodium dodecyl (lauryl) sulfate (SDS)-NaOH protocol. Of these, 94% were recovered with the BacT/Alert 3D system (Organon Teknika, Durham, N.C.) and 79% were recovered on Löwenstein-Jensen (LJ) medium. Mean times to detection of organisms of the Mycobacterium tuberculosis complex (n = 47) were 2...
متن کاملMulticenter evaluation of the MB/BACT system for susceptibility testing of Mycobacterium tuberculosis.
The reliability of the MB/BACT system for susceptibility testing of Mycobacterium tuberculosis to pyrazinamide, rifampin, isoniazid, streptomycin, and ethambutol was compared to the BACTEC 460TB system. The proportion method was used to resolve discrepant results by an independent arbiter. Two interpretative methods were used, with an undiluted control (direct control) and a diluted control (10...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of clinical microbiology
دوره 45 1 شماره
صفحات -
تاریخ انتشار 2007