Long-term performance of the St Jude Riata 1580-1582 ICD lead family.

نویسندگان

  • S D A Valk
  • D A M J Theuns
  • L Jordaens
چکیده

OBJECTIVE Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580-1582 were studied, retrospectively, before the official recall. METHODS All 374 patients at the Erasmus Medical Center between July 2003 and December 2007 with a 1580, 1581 or 1582 shock lead. RESULTS The majority of the patients were male (78 %), with a median age of 60 years (IQR 52-70); primary prevention in 61 %. Median follow-up was 60.3 months (IQR 35.5-73.2), with 117 (31 %) patients dying. Electrical abnormalities (mainly noise, 65 %) were observed in 20/257 patients (7.8 %). Definite conductor externalisation was confirmed with fluoroscopy or chest X-ray in 16 patients, and in one after extraction. One patient presented with a drop in the high-voltage impedance trend with a short circuit of the ICD system during defibrillation testing, and needed to be shocked externally. In 8 more patients, conductor externalisation was found during an elective procedure. No predictors of externalisation could be found, except for the use of single coil (p = 0.02). Median time to conductor externalisation was 5 years (IQR 3.1-6.2). Lead externalisation was observed in 5.4 % (95 % CI 3.1-9.3) at 5 years and 22.7 % (95 % CI 13.6-36.6) at 8 years. CONCLUSION A high incidence of insulation defects associated with conductor externalisation in the Riata ICD lead family is observed. The mode of presentation is diverse. This type of insulation failure can lead to failure of therapy delivery.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transvenous Implantable Cardioverter‐Defibrillator (ICD) Lead Performance: A Meta‐Analysis of Observational Studies

BACKGROUND Despite the widespread use of implantable cardioverter-defibrillators (ICDs) in clinical practice, concerns exist regarding ICD lead durability. The performance of specific lead designs and factors determining this in large populations need clarification. METHODS AND RESULTS The Medline, Embase, and Cochrane Collaboration databases were searched for studies including ≥2 of the most...

متن کامل

Thrombus formation on a defibrillator lead with conductor externalization

A 51 year old man with arrhythmogenic right ventricular cardiomyopathy reported non-specific symptoms of muscle aches and shivering at outpatient review. A dual chamber implantable cardioverter defibrillator (ICD) had been inserted seven years previously. Fluoroscopic lead screening had previously demonstrated conductor externalisation (Figure and Video) of a single coil Riata ICD lead (model n...

متن کامل

A case of Durata ICD lead coil externalization: Inside-out lead abrasion?

The St Jude Medical Durata lead was produced in response to the recalled Riata and Riata ST leads in November 2011 after concerns of lead durability and inside-out abrasion. In order to protect against this susceptibility, the Riata ST Optim and Durata leads were coated with an abrasionresistant layer of silicone–polyurethane copolymer (Optim) tubing. We report a case of Durata lead extraction ...

متن کامل

A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation

A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead...

متن کامل

Canadian Registry of Implantable Electronic Device Outcomes

An advisory related to the Riata and Riata ST leads was announced on November 28, 2011 by St. Jude Medical after these leads had been removed from the market the year before. This advisory described the possibility of conductors within the implantable cardioverter–defibrillator (ICD) lead becoming externalized because of an inside-out abrasion mechanism. This is a novel failure mechanism, not r...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation

دوره 21 3  شماره 

صفحات  -

تاریخ انتشار 2013