Validation of the integration of technology that measures additional “vascular” indices into an ambulatory blood pressure monitoring system
نویسندگان
چکیده
BACKGROUND The objective of this study was to validate the novel integration of oscillometric (Vasotens(®)) technology into a BPLab(®) ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method. METHODS The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional "classic" SphygmoCor device. RESULTS Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of -1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of -2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV). CONCLUSION The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs.
منابع مشابه
A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Blood pressure monitoring based on pulse transit or arrival time has been the focus of much research in order to design ambulatory blood pressure monitors. The accuracy of these monitors is limited by several challenges, such as acquisition and processing of physiological signals as well as changes in vascular tone and the pre-ejection period. In this work, a literature survey covering recent d...
متن کاملPhysiological Measurement TOPICAL REVIEW A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Blood pressure monitoring based on pulse transit or arrival time has been the focus of much research in order to design ambulatory blood pressure monitors. The accuracy of these monitors is limited by several challenges, such as acquisition and processing of physiological signals as well as changes in vascular tone and the pre-ejection period. In this work, a literature survey covering recent d...
متن کاملA survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Blood pressure monitoring based on pulse transit or arrival time has been the focus of much research in order to design ambulatory blood pressure monitors. The accuracy of these monitors is limited by several challenges, such as acquisition and processing of physiological signals as well as changes in vascular tone and the pre-ejection period. In this work, a literature survey covering recent d...
متن کاملA survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Blood pressure monitoring based on pulse transit or arrival time has been the focus of much research in order to design ambulatory blood pressure monitors. The accuracy of these monitors is limited by several challenges, such as acquisition and processing of physiological signals as well as changes in vascular tone and the pre-ejection period. In this work, a literature survey covering recent d...
متن کاملA survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Blood pressure monitoring based on pulse transit or arrival time has been the focus of much research in order to design ambulatory blood pressure monitors. The accuracy of these monitors is limited by several challenges, such as acquisition and processing of physiological signals as well as changes in vascular tone and the pre-ejection period. In this work, a literature survey covering recent d...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 7 شماره
صفحات -
تاریخ انتشار 2014