Transcatheter closure of perimembranous ventricular septal defects with the Amplatzer asymmetric ventricular septal defect occluder: preliminary experience in children.

نویسندگان

  • B D Thanopoulos
  • G S Tsaousis
  • E Karanasios
  • N G Eleftherakis
  • C Paphitis
چکیده

OBJECTIVE To close perimembranous ventricular septal defects (PMVSDs) in children with the new Amplatzer asymmetric ventricular septal defect occluder (AAVSDO). PATIENTS AND DESIGN 10 children, aged 1.5-12 years, with PMVSDs underwent transcatheter closure with the AAVSDO. The device consists of two low profile disks made of Nitinol wire mesh with a 1.5 mm connecting waist. The left disk is 5 mm towards the apex and only 0.5 mm towards the aortic valve. The right disk is 4 mm larger than the waist. The prosthesis diameter was chosen to be 1-2 mm larger than the largest diameter of the defect (determined by transoesophageal echocardiography and angled angiocardiography). A 7-8 French gauge sheath was used to deliver the AAVSDO. RESULTS The PMVSD diameter ranged from 2-8 mm. The device diameter ranged from 4-8 mm. After deployment of the prosthesis there was no residual shunt in 9 of 10 patients (90%). In one patient there was a trivial residual shunt that disappeared at the three month follow up. Three patients developed transient complete left bundle branch block. No other complications were observed. CONCLUSIONS The AAVSDO appears to be a promising device for transcatheter closure of PMVSDs in children. Further studies are required to document its efficacy, safety, and long term results in a larger patient population.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

CONGENITAL HEART DISEASE Transcatheter closure of perimembranous ventricular septal defects with the Amplatzer asymmetric ventricular septal defect occluder: preliminary experience in children

Objective: To close perimembranous ventricular septal defects (PMVSDs) in children with the new Amplatzer asymmetric ventricular septal defect occluder (AAVSDO). Patients and design: 10 children, aged 1.5–12 years, with PMVSDs underwent transcatheter closure with the AAVSDO. The device consists of two low profile disks made of Nitinol wire mesh with a 1.5 mm connecting waist. The left disk is 5...

متن کامل

Transcatheter closure of iatrogenic perimembranous ventricular septal defect after aortic valve and ascending aorta replacement using an Amplatzer membranous ventricular septal occluder.

Iatrogenic perimembranous ventricular septal defect is a rare complication after surgical replacement of the aortic valve, and so transcatheter closure of such a defect is not a routine procedure. We report the successful closure of an iatrogenic perimembranous ventricular septal defect which occurred after the replacement of the aortic valve and ascending aorta.

متن کامل

Transcatheter closure of perimembranous ventricular septal defects with Amplatzer occluder assessed by real-time three-dimensional echocardiography.

We report our initial experience in two patients using real-time three-dimensional echocardiography to assess perimembranous ventricular septal defect and device morphology and their relation with contiguous cardiac structure. Defect size and rims as well as device position and profile were displayed from the three-dimensional "en face" views. We think that real-time three-dimensional echocardi...

متن کامل

Surgical repair of tricuspid valve leaflet tear following percutaneous closure of perimembranous ventricular septal defect using Amplatzer duct occluder I: Report of two cases

Tricuspid valve (TV) injury following transcatheter closure of perimembranous ventricular septal defect (PMVSD) with Amplatzer ductal occluder I (ADO I), requiring surgical repair, is rare. We report two cases of TV tear involving the anterior and septal leaflets following PMVSD closure using ADO I. In both the patients, the subvalvular apparatus remained unaffected. The patients presented with...

متن کامل

Efficacy and Safety of Using Amplatzer Ductal Occluder for Transcatheter Closure of Perimembranous Ventricular Septal Defect in Pediatrics

BACKGROUND Perimembranous Ventricular Septal Defect (PMVSD) is the most common subtype of ventricular septal defects. Transcatheter closure of PMVSD is a challenging procedure in management of moderate or large defects. OBJECTIVES The purpose of this study was to show that transcatheter closure of perimembranous ventricular septal defect with Amplatzer Ductal Occluder (ADO) is an effective an...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Heart

دوره 89 8  شماره 

صفحات  -

تاریخ انتشار 2003