OBJECTIVES This study classified and quantified the variation in fractional flow reserve (FFR) due to fluctuations in systemic and coronary hemodynamics during intravenous adenosine infusion. BACKGROUND Although FFR has become a key invasive tool to guide treatment, questions remain regarding its repeatability and stability during intravenous adenosine infusion because of systemic effects that can alter driving pressure and heart rate. METHODS We reanalyzed data from the VERIFY (VERification of Instantaneous Wave-Free Ratio and Fractional Flow Reserve for the Assessment of Coronary Artery Stenosis Severity in EverydaY Practice) study, which enrolled consecutive patients who were infused with intravenous adenosine at 140 μg/kg/min and measured FFR twice. Raw phasic pressure tracings from the aorta (Pa) and distal coronary artery (Pd) were transformed into moving averages of Pd/Pa. Visual analysis grouped Pd/Pa curves into patterns of similar response. Quantitative analysis of the Pd/Pa curves identified the "smart minimum" FFR using a novel algorithm, which was compared with human core laboratory analysis. RESULTS A total of 190 complete pairs came from 206 patients after exclusions. Visual analysis revealed 3 Pd/Pa patterns: "classic" (sigmoid) in 57%, "humped" (sigmoid with superimposed bumps of varying height) in 39%, and "unusual" (no pattern) in 4%. The Pd/Pa pattern repeated itself in 67% of patient pairs. Despite variability of Pd/Pa during the hyperemic period, the "smart minimum" FFR demonstrated excellent repeatability (bias -0.001, SD 0.018, paired p = 0.93, r(2) = 98.2%, coefficient of variation = 2.5%). Our algorithm produced FFR values not significantly different from human core laboratory analysis (paired p = 0.43 vs. VERIFY; p = 0.34 vs. RESOLVE). CONCLUSIONS Intravenous adenosine produced 3 general patterns of Pd/Pa response, with associated variability in aortic and coronary pressure and heart rate during the hyperemic period. Nevertheless, FFR - when chosen appropriately - proved to be a highly reproducible value. Therefore, operators can confidently select the "smart minimum" FFR for patient care. Our results suggest that this selection process can be automated, yet comparable to human core laboratory analysis.
منابع مشابه
Role of fractional and coronary flow reserve in clinical decision making in intermediate coronary lesions
Adequate patient selection for percutaneous coronary intervention is of the greatest importance in order to minimize early and late complications. Therefore, objective evidence for myocardial ischemia is mandatory for the management of patients with coronary artery disease, in particular in multivessel disease and those with intermediate lesions (40–70% diameter stenosis on angiography). The us...
متن کاملT2-prepared steady-state free-precession for detection of hemodynamic significance of coronary artery stenosis - a comparison to fractional flow reserve
Background BOLD CMR has been shown to be able to detect myocardial perfusion differences. However, validation of BOLD CMR against FFR is lacking. Aim of our study was to analyze the potential diagnostic accuracy of blood oxygen level-dependent (BOLD) cardiac magnetic resonance imaging (CMR) in comparison to invasively measured fractional flow reserve (FFR) which served as gold standard for a he...
متن کاملComplex Coronary Hemodynamics - Simple Analog Modelling as an Educational Tool
OBJECTIVE Invasive coronary angiography remains the cornerstone for evaluation of coronary stenoses despite there being a poor correlation between luminal loss assessment by coronary luminography and myocardial ischemia. This is especially true for coronary lesions deemed moderate by visual assessment. Coronary pressure-derived fractional flow reserve (FFR) has emerged as the gold standard for ...
متن کاملFractional Flow Reserve Assessment of a Significant Coronary Stenosis Masked by a Downstream Serial Lesion
Fractional flow reserve (FFR) has been recognized as an effective tool to determine functional significance in intermediate coronary lesions and FFR-guided percutaneous coronary intervention (PCI) improves clinical outcomes. However, hemodynamic interaction between serial stenoses within one coronary artery complicates the assessment of functional severity of each individual lesion. We present ...
متن کاملNoninvasive Fractional Flow Reserve Derived From Coronary CT Angiography: Clinical Data and Scientific Principles.
Fractional flow reserve derived from coronary computed tomography angiography enables noninvasive assessment of the hemodynamic significance of coronary artery lesions and coupling of the anatomic severity of a coronary stenosis with its physiological effects. Since its initial demonstration of feasibility of use in humans in 2011, a significant body of clinical evidence has developed to evalua...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- JACC. Cardiovascular interventions
دوره 8 8 شماره
صفحات -
تاریخ انتشار 2015