Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement.

نویسندگان

  • James Pittman
  • Shahar Bar-Yosef
  • John SumPing
  • Matthew Sherwood
  • Jonathan Mark
چکیده

OBJECTIVE Pulse contour analysis can be used to provide beat-to-beat cardiac output (CO) measurement. The current study sought to evaluate this technique by comparing its results with lithium dilution CO (LiCO) measurements. DESIGN Prospective, observational study. SETTING Surgical intensive care unit. PATIENTS Twenty-two patients after cardiac or major noncardiac surgery. MEASUREMENTS After initial calibration of the pulse contour CO (PCO) method, CO was measured by PCO and by LiCO methods at 4, 8, 16, and 24 hrs. Recalibration of PCO was performed every 8 hrs. The systemic vascular resistance and dynamic response characteristics of the arterial catheter-transducer system were measured at each time point to determine whether these influenced the agreement between PCO and LiCO methods. MAIN RESULTS There was an excellent correlation between methods (r = .94). Bias was small (-0.005 L/min), and clinically acceptable limits of agreement were demonstrated between techniques. Although many catheter-transducer systems had poor dynamic response characteristics, this did not influence the level of agreement between the two techniques. An increase in systemic vascular resistance between two time points did tend to cause overestimation of LiCO by the PCO. CONCLUSIONS PCO measurement compared well with the lithium dilution method and can be considered an accurate technique for measuring beat-to-beat CO with limited risk to the patient.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output.

A new indicator dilution technique for measurement of cardiac output is described. Lithium chloride is injected via a central venous catheter and its dilution curve measured in arterial blood using a lithium-selective electrode. We assessed the lithium dilution cardiac output measurement (LiDCO) and a conventional thermodilution cardiac output measurement (ThDCO) by comparing the results of bot...

متن کامل

Evaluation of the Pulse - Contour Method of Determining Stroke

one patients with ischemic heart disease and 21 patients with mitral valve disease were subjected to a variety of hemodynamic interventions. These included exercise, isoproterenol infusion, and practolol, morphine or pentazocine administration. The pulse-contour estimations, using three different formulae derived by Warner, Kouchoukos, and Herd, were compared with indicator-dilution outputs. Th...

متن کامل

Evaluation of the Pulse - Contour Method of Determining Stroke Volume in Man By EDWIN

one patients with ischemic heart disease and 21 patients with mitral valve disease were subjected to a variety of hemodynamic interventions. These included exercise, isoproterenol infusion, and practolol, morphine or pentazocine administration. The pulse-contour estimations, using three different formulae derived by Warner, Kouchoukos, and Herd, were compared with indicator-dilution outputs. Th...

متن کامل

Evaluation of the pulse-contour method of determining stroke volume in man.

one patients with ischemic heart disease and 21 patients with mitral valve disease were subjected to a variety of hemodynamic interventions. These included exercise, isoproterenol infusion, and practolol, morphine or pentazocine administration. The pulse-contour estimations, using three different formulae derived by Warner, Kouchoukos, and Herd, were compared with indicator-dilution outputs. Th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Critical care medicine

دوره 33 9  شماره 

صفحات  -

تاریخ انتشار 2005