Behavioral Patterns and Associations with Glucose Control During 12-Week Randomized Free-Living Clinical Trial of Day and Night Hybrid Closed-Loop Insulin Delivery in Adults with Type 1 Diabetes

نویسندگان

  • Ali Emami
  • Malgorzata E. Willinska
  • Hood Thabit
  • Lalantha Leelarathna
  • Sara Hartnell
  • Sibylle Dellweg
  • Carsten Benesch
  • Julia K. Mader
  • Manuel Holzer
  • Harald Kojzar
  • Thomas R. Pieber
  • Sabine Arnolds
  • Mark L. Evans
  • Roman Hovorka
چکیده

OBJECTIVES We evaluated patterns of meal intake, insulin bolus delivery, and fingerstick glucose measurements during hybrid closed-loop and sensor-augmented pump (SAP) therapy, including associations with glucose control. METHODS Data were retrospectively analyzed from pump-treated adults with type 1 diabetes who underwent, in random order, 12 weeks free-living closed-loop (n = 32) and 12 weeks SAP (n = 33) periods. We quantified daily patterns of main meals, snacks, prandial insulin boluses, correction boluses, and fingerstick glucose measurements by analyzing data recorded on the study glucometer and on study insulin pump. RESULTS We analyzed 1942 closed-loop days and 2530 SAP days. The total number of insulin boluses was reduced during closed-loop versus SAP periods by mean 1.0 per day (95% confidence interval 0.6-1.4, P < 0.001) mainly because of a reduced number of correction boluses by mean 0.7 per day (0.4-1.0, P < 0.001). Other behavioral patterns were unchanged. The carbohydrate content of snacks but not the number of snacks was positively correlated with (1) glycemic variability as measured by standard deviation of sensor glucose (closed-loop P < 0.05; SAP P < 0.01), (2) mean sensor glucose (P < 0.05), and (3) postintervention HbA1c (P < 0.05). Behavioral patterns explained 47% of between-subject variance in glucose variability during SAP period and 30%-33% of variance of means sensor glucose and postintervention HbA1c. CONCLUSION Fewer correction boluses are delivered during closed-loop period. The size of snacks appears to worsen glucose control possibly because of carbohydrate-rich content of snacks. Modifiable behavioral patterns may be important determinants of glucose control.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Home Use of Day-and-Night Hybrid Closed-Loop Insulin Delivery in Suboptimally Controlled Adolescents With Type 1 Diabetes: A 3-Week, Free-Living, Randomized Crossover Trial

OBJECTIVE This study evaluated the feasibility, safety, and efficacy of day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes under free-living conditions. RESEARCH DESIGN AND METHODS In an open-label randomized crossover study, 12 suboptimally controlled adolescents on insulin pump therapy (mean ± SD age 14.6 ± 3.1 years; HbA1c 69 ± 8 mmol/mol [8.5 ± 0.7%]; du...

متن کامل

Day-and-Night Hybrid Closed-Loop Insulin Delivery in Adolescents With Type 1 Diabetes: A Free-Living, Randomized Clinical Trial

OBJECTIVE To evaluate feasibility, safety, and efficacy of day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes under free-living conditions without remote monitoring or supervision. RESEARCH DESIGN AND METHODS In an open-label, randomized, free-living, crossover study design, 12 adolescents receiving insulin pump therapy (mean [±SD] age 15.4 ± 2.6 years; HbA1...

متن کامل

Day-and-night glycaemic control with closed-loop insulin delivery versus conventional insulin pump therapy in free-living adults with well controlled type 1 diabetes: an open-label, randomised, crossover study

BACKGROUND Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve glucose control while alleviating the risk of hypoglycaemia in adults with HbA1c below 7·5% (58 mmol/mol). METHODS In this open-label, randomised, crossover study, we recruited adults ...

متن کامل

Overnight Closed-Loop Insulin Delivery in Young People With Type 1 Diabetes: A Free-Living, Randomized Clinical Trial

OBJECTIVE To evaluate feasibility, safety, and efficacy of overnight closed-loop insulin delivery in free-living youth with type 1 diabetes. RESEARCH DESIGN AND METHODS Overnight closed loop was evaluated at home by 16 pump-treated adolescents with type 1 diabetes aged 12-18 years. Over a 3-week period, overnight insulin delivery was directed by a closed-loop system, and on another 3-week per...

متن کامل

Day and night home closed-loop insulin delivery in adults with type 1 diabetes: three-center randomized crossover study.

OBJECTIVE To evaluate the feasibility of day and night closed-loop insulin delivery in adults with type 1 diabetes under free-living conditions. RESEARCH DESIGN AND METHODS Seventeen adults with type 1 diabetes on insulin pump therapy (means ± SD age 34 ± 9 years, HbA1c 7.6 ± 0.8%, and duration of diabetes 19 ± 9 years) participated in an open-label multinational three-center crossover study....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 19  شماره 

صفحات  -

تاریخ انتشار 2017