Calculation of SAR for Transmit Coil Arrays.
نویسندگان
چکیده
Transmit coil arrays allowing independent control of individual coil drives facilitate adjustment of the B(1) field distribution, but when the B(1) field distribution is changed the electric field and SAR distributions are also altered. This makes safety evaluation of the transmit array a challenging problem because there are potentially an infinite number of possible field distributions in the sample. Local SAR levels can be estimated with numerical calculations, but it is not practical to perform separate full numerical calculations for every current distribution of interest. Here we evaluate superposition of separate electric field calculations-one for each coil-for predicting SAR in a full numerical calculation where all coils are driven simultaneously. It is important to perform such an evaluation because the effects of coil coupling may alter the result. It is shown that while there is good agreement between the superimposed and simultaneous drive results when using current sources in the simulations, the agreement is not as good when voltage sources are used. Finally, we compare maximum local SAR levels for B(1) field distributions that are either unshimmed or shimmed over one of three regions of interest. When B(1) field homogeneity is improved in a small region of interest without regard for SAR, the maximum local SAR can become very high.
منابع مشابه
Comparison of simulated parallel transmit body arrays at 3 T using excitation uniformity, global SAR, local SAR, and power efficiency metrics.
PURPOSE We compare the performance of eight parallel transmit (pTx) body arrays with up to 32 channels and a standard birdcage design. Excitation uniformity, local specific absorption rate (SAR), global SAR, and power metrics are analyzed in the torso at 3 T for radiofrequency (RF)-shimming and 2-spoke excitations. METHODS We used a fast cosimulation strategy for field calculation in the pres...
متن کاملFour Channel Transceiver Array for Functional Magnetic Resonance Spectroscopy in the Human Visual Cortex at 9.4 T
Introduction: Functional magnetic resonance spectroscopy (fMRS) at ultra-high field (UHF) strength (≥7T) is a method to investigate the metabolism of the human brain during task induced stimulation [1] or pharmalogical challenge [2]. A region of special interest is the visual cortex which undergoes metabolic changes during visual stimulation [1]. The UHF thereby allows for increased spectral an...
متن کاملCurrent Controlled Transmit/Receive Elements in Parallel Excitation and Parallel Imaging
Introduction and Motivation State-of-the-art MRI typically relies on a power driven body coil and a receive-coil array for parallel imaging (PI). However, at 3T and above parallel excitation (PEX) techniques utilizing multi element transmit coil arrays are much better suited to handle B1-homogeneity issues. In power driven transmit arrays, electromagnetic coupling between the coil elements pose...
متن کاملZ-stacked RF array design enhances parallel transmit multiband RF performance in whole brain simultaneous multislice imaging at 7T
Introduction: Simultaneous MultiSlice (SMS) MR imaging (1) using MultiBand (MB) RF pulses is becoming increasingly popular in the neuroimaging community (2-4). Recently, there has been an interest in utilizing multielement RF arrays in combination with multichannel (pTx) MB pulse design to reduce transmit B1 (B1+) inhomogeneity and SAR (5,6) for SMS/MB imaging. Meanwhile, it has been shown that...
متن کاملTransmit Arrays and Circuitry
For body dimensions at 3T and head dimensions at 7T and above the RF wavelength in dielectric tissue is comparable or smaller than the dimensions of the human anatomy. This in turn leads to RF phase related traveling time differences, prominent wave behavior and the potential for a significant difference between transmit B1 + and receive B1 fields [5-10]. It is possible to correct some of the r...