Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis.

نویسندگان

  • S G Sakka
  • J Kozieras
  • O Thuemer
  • N van Hout
چکیده

BACKGROUND Recently, continuous monitoring of cardiac output (CO) based on pulse contour analysis (Vigileo) has been introduced into practice. In this clinical study, we evaluated the accuracy of this system by comparing it with the transpulmonary thermodilution technique (TPID) in septic patients. METHODS We studied 24 mechanically ventilated patients with septic shock (16 male, 8 female, age 26-77 yr) receiving treatment with norepinephrine who for clinical indication underwent haemodynamic monitoring by the transpulmonary thermodilution technique using a PiCCO plus system (Pulsion Medical Systems, Munich, Germany). In parallel, arterial pulse contour was applied using the femoral arterial pressure curve (FloTrac pressure sensor, Vigileo monitor, Edwards Lifesciences, Irvine, USA). After baseline measurement, mean arterial pressure was elevated by increasing norepinephrine dosage, and CO was measured again before mean arterial pressure was reduced back to baseline levels. Fluid status and ventilator settings remained unchanged throughout. At each time point, CO by transpulmonary thermodilution was calculated from three central venous bolus injections of 15 ml of saline (<8 degrees C). Linear regression and the Bland-Altman method were used for statistical analysis. RESULTS Overall, CO was 6.7 (sd 1.8) (3.2-10.1) litre min(-1) for CO(TPID) and 6.2 (2.4) (3.0-17.6) litre min(-1) for CO(Vigileo((R))). Linear regression revealed: CO(Vigileo) = 1.54 + 0.72 x CO(TPID) litre min(-1), r(2) = 0.26 (P < 0.0001). Mean bias between techniques [CO(TPID)-CO(Vigileo)] was 0.5 litre min(-1) (SD 2.3 litre min(-1)). Correlation coefficients at the three time points were not significantly different from each other. CONCLUSIONS Pulse contour analysis-derived CO (Vigileo system) underestimates CO(TPID) and is not as reliable as transpulmonary thermodilution in septic patients.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Comparison of Third-Generation Semi-Invasive Arterial Waveform Analysis with Thermodilution in Patients Undergoing Coronary Surgery

Uncalibrated semi-invasive continous monitoring of cardiac index (CI) has recently gained increasing interest. The aim of the present study was to compare the accuracy of CI determination based on arterial waveform analysis with transpulmonary thermodilution. Fifty patients scheduled for elective coronary surgery were studied after induction of anaesthesia and before and after cardiopulmonary b...

متن کامل

Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery

INTRODUCTION Uncalibrated arterial pulse power analysis has been recently introduced for continuous monitoring of cardiac index (CI). The aim of the present study was to compare the accuracy of arterial pulse power analysis with intermittent transpulmonary thermodilution (TPTD) before and after cardiopulmonary bypass (CPB). METHODS Forty-two patients scheduled for elective coronary surgery we...

متن کامل

Accuracy of Cardiac Output by Nine Different Pulse Contour Algorithms in Cardiac Surgery Patients: A Comparison with Transpulmonary Thermodilution

Objective. Today, there exist several different pulse contour algorithms for calculation of cardiac output (CO). The aim of the present study was to compare the accuracy of nine different pulse contour algorithms with transpulmonary thermodilution before and after cardiopulmonary bypass (CPB). Methods. Thirty patients scheduled for elective coronary surgery were studied before and after CPB. A ...

متن کامل

Validation of cardiac output monitoring based on uncalibrated pulse contour analysis vs transpulmonary thermodilution during off-pump coronary artery bypass grafting.

BACKGROUND Cardiac output monitoring, as a part of a goal-directed haemodynamic management, has been shown to improve perioperative outcome in high-risk patients undergoing major surgical interventions. However, thorough validation of cardiac output monitoring devices in different clinical conditions is warranted. The aim of our study was to compare the reliability of a novel system for cardiac...

متن کامل

CARDIOVASCULAR Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients

Editor’s key points † Previously, continuous cardiac output (CO) measurement techniques required the insertion of a pulmonary artery catheter. † The PiCCO and VolumeView systems use arterial pulse contour analysis for continuous CO measurement. † Both require calibration with transpulmonary thermodilution (TPTD) measurements using a central venous catheter. † The current study compared the perf...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 99 3  شماره 

صفحات  -

تاریخ انتشار 2007