Tissue rupture and bubble formation during radiofrequency catheter ablation: "echoes of a pop".
نویسندگان
چکیده
The popping phenomenon during radiofrequency catheter ablation is well known to experienced electrophysiologists. When the electrode-tissue interface temperature increases rapidly above the boiling point, it can vaporize blood, thus causing a mini-explosion and an audible pop. Evaporation may occur intramurally, leading to gas bubble formation within the tissue. With continued energy application, this bubble expands and erupts through the weakest path, cleaving the tissues. Whether the intramural vaporization and explosion can lead to shock-wave production, with consequent cavitation and microbubble formation akin to high-energy direct-current ablation, is not known. An intracardiac echogram at 7.5 MHz using the AcuNav catheter (Acuson Inc) during an experimental study on a pig model is shown in Figure 1. Linear epicardial radiofrequency energy applications to the left atrial appendage were delivered under direct vision using an irrigated tip catheter. Serial 60-s lesions at a power output of 25 W were applied to create a linear lesion. During one of the applications, there was an audible pop, which was seen as a mini-explosion of microbubbles in the left atrial cavity. Perforation of the left atrium occurred at this site (note the forceps through the hole in Figure 2). Popping during radiofrequency catheter ablation is associated with unwanted effects such as blood boiling and vaporization, endocardial rupture, clot formation over the catheter/ endocardium, catheter dislocation, and impedance rise. Poor electrode-tissue contact with high power during energy delivery increases the probability of pops. The images presented are a visual testimony to these physical effects within the heart; in this case, these effects were of such severity that they perforated the atrial wall.
منابع مشابه
Giant J-wave (Osborn wave) unrelated to hypothermia.
contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 2008;1:354–62. 3. Shah DC, Leo G, Aeby N, Le Royer E, Sunthorn H, Burri H et al. Evaluation of a new catheter sensor for real-time measurement of tissue contact. Heart Rhythm 2006;3:S75–6. Abstract. 4. Ikeda A, Nakagawa H, Sha...
متن کاملAn Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation
Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...
متن کاملNovel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus.
BACKGROUND An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation...
متن کاملEffect of Left Atrial Ablation Process and Strategy on Microemboli Formation During Irrigated Radiofrequency Catheter Ablation in an In Vivo Model.
BACKGROUND Formation of microemboli during catheter ablation has been suggested as a cause for asymptomatic cerebral emboli. However, it is unknown which part of the process and ablation setting/strategy is most strongly related to this occurrence. METHODS AND RESULTS A total of 27 pigs were used. Catheter/sheath manipulations in left atrium were performed in 25 of 27 pigs outfitted with micr...
متن کاملPulmonary Embolism as a Complication after radiofrequency catheter ablation Study: A Case Report
Pulmonary thromboembolism (PTE) is a fatal condition that may rarely occur due to complications of coronary catheter insertion. In this case report, a 41-year-old male was presented 48 hours after radiofrequency catheter ablation(RFCA) for the management of Wolf-Parkinson-White syndromewith acute onset of dyspnea, hemoptysis, and chest pain. The physical examination revealed coarse crackles in ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Circulation
دوره 103 9 شماره
صفحات -
تاریخ انتشار 2001