Development of and Clinical Experience with a Simple Device for Performing Intraoperative Fluorescein Fluorescence Cerebral Angiography: Technical Notes
نویسندگان
چکیده
To perform intraoperative fluorescence angiography (FAG) under a microscope without an integrated FAG function with reasonable cost and sufficient quality for evaluation, we made a small and easy to use device for fluorescein FAG (FAG filter). We investigated the practical use of this FAG filter during aneurysm surgery, revascularization surgery, and brain tumor surgery. The FAG filter consists of two types of filters: an excitatory filter and a barrier filter. The excitatory filter excludes all wavelengths except for blue light and the barrier filter passes long waves except for blue light. By adding this FAG filter to a microscope without an integrated FAG function, light from the microscope illuminating the surgical field becomes blue, which is blocked by the barrier filter. We put the FAG filter on the objective lens of the operating microscope correctly and fluorescein sodium was injected intravenously or intra-arterially. Fluorescence (green light) from vessels in the surgical field and the dyed tumor were clearly observed through the microscope and recorded by a memory device. This method was easy and could be performed in a short time (about 10 seconds). Blood flow of small vessels deep in the surgical field could be observed. Blood flow stagnation could be evaluated. However, images from this method were inferior to those obtained by currently commercially available microscopes with an integrated FAG function. In brain tumor surgery, a stained tumor on the brain surface could be observed using this method. FAG could be performed with a microscope without an integrated FAG function easily with only this FAG filter.
منابع مشابه
Intra-Arterial Fluorescence Angiography with Injection of Fluorescein Sodium from the Superficial Temporal Artery during Aneurysm Surgery: Technical Notes
Intra-arterial fluorescence angiography from a catheter inserted into the external carotid artery (ECA) via the superficial temporal artery (STA) allowed us to satisfactorily evaluate cerebral arterial and venous blood flow. We report this novel method that allowed for repeated angiography within minutes with a low risk of complications due to catheter placement from the STA. The STA was secure...
متن کاملIn-House Handmade Camera for Indocyanine Green Video Angiography
Background and Aim: Vascular imaging during surgical procedures is very important and has many applications. There are several methods for intraoperative vascular assessment such as intraoperative angiography, Doppler, and fluorescence-based techniques. The latter group and specially the Indocyanine Green (ICG) Video Angiography (VA) is commonly used for vascular surgery and sentinel node...
متن کاملIntraoperative Near - infrared Indocyanine Green Videoangiography
Fluorescence angiography was first used by ophthalmologists to measure retinal blood flow by using the fluorescent dye fluorescein. Feindel et al.1,2 were the first to apply the concept of fluorescence angiography to the intraoperative visualisation of cerebral vertebral arteries and cerebral microcirculation in patients undergoing neurosurgical procedures.3,4 With the use of indocyanine green ...
متن کاملTransradial Neurointerventional Approach to Basilar Tip Aneurysm: Case Report and Technical Note
Cerebral angiography through transfemoral approach is preferred in most of the situation though in cardiovascular surgery transradial angiography is accounted as the superior standard. The present study aimed at presenting our early experience of transradial approach to cerebral angiography and intervention of a patient with basilar tip aneurysm that had inappropriate vasculature to perform rou...
متن کاملFeasibility and Utility of Transradial Cerebral Angiograpy: Experience during the Learning Period
OBJECTIVE We wanted to present our experiences for performing transradial cerebral angiography during the learning period, and we also wanted to demonstrate this procedure's technical feasibility and utility in various clinical situations. MATERIALS AND METHODS Thirty-two patients were enrolled in the study. All of them had unfavorable situations for performing transfemoral angiography, i.e.,...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 56 شماره
صفحات -