Autoinflation of saline-filled inflatable breast implants.
نویسنده
چکیده
Spontaneous autoinflation of saline-filled breast implants is a rare phenomenon; only 20 cases have been reported in the world literature. Over the past seven years, three patients have presented with significant unilateral autoinflation of their smooth, single-lumen, round, saline-filled implants. This developed at various times: progressively over 23 years with a Simaplast implant; between the ninth and 10th year after augmentation with a leaflet valve implant; and slowly over the first four years with a leaflet valve implant. The etiology of the autoinflation was shown to be different for the two types of implants. The Simaplast implant had likely been injected with a hypertonic filling solution - one that was twice as concentrated as 'normal saline'. This would have created an osmotic gradient, which would have facilitated autoexpansion by diffusion. The implant solution remained clear and transparent. In addition, there were no detectable levels of glucose, uric acid or albumin in the fluid. By contrast, auto-inflation of the leaflet valve implants likely resulted from mechanical alterations of the valve mechanism. This type of implant is known to have a high deflation rate, with frequent partial deflations. It is interesting that one of the patients with the leaflet valve implants presented with an autoexpansion on one side and a partial deflation on the other side. Both implants were from the same lot number. The same mechanism that caused partial deflations may have also allowed fluid from the implant pocket to pass through the valve into the lumen of these implants. This could allow glucose, protein and cellular elements to enter into the lumen (these would not pass through an intact elastomeric shell) which would create an osmotic gradient, allowing water to enter the elastomeric shell by diffusion. The fluid in these leaflet valve implants was brownish yellow, very viscous and turbid. It contained elevated levels of glucose and uric acid which would not have passed through the elastomeric shell. Over the past 10 years, four different theories have been proposed in an attempt to explain the etiology of autoinflation. However, the findings of the present study indicate that there are only two mechanisms - a hypertonic filling solution and alterations of the valve mechanism.
منابع مشابه
Failure properties of leaf valve inflatable saline breast implants
In 1963, Cronin and Gerow (1) introduced the Silastic (silicone gel, Dow Corning, Tennessee) breast implant, opening the new era of breast augmentation surgery. This original ‘Cronin’ implant was made of a thick gel contained in a thick, rigid, elastomeric envelope, which had a Dacron (Dupont, Delaware) backing on the posterior surface to anchor the implant to the chest wall to attempt to reduc...
متن کاملExperience of Breast Augmentation in Pakistani Females
BACKGROUND Breast augmentation is an elective surgery used to increase the size of the breast in females. This study was undertaken to determine the breast augmentation in Pakistani females. METHODS From 2006 to 2011, 43 female patients who underwent breast augmentation via infra-mammary, periareolar, trans-umbilical and fat transfer were enrolled. For augmentation, saline-filled or gel-fill...
متن کاملProspective Analysis of Primary Breast Augmentation on Body Image Using the BREAST-Q: Results from a Nationwide Study
BACKGROUND The Breast Implant Follow-up Study is a large, ongoing observational study of women who received Natrelle round silicone-filled or saline-filled breast implants. This analysis describes patient-reported outcomes in the cohort who underwent breast augmentation. METHODS Subjects prospectively completed two validated scales of the BREAST-Q (satisfaction with breasts and psychosocial w...
متن کاملErosion of an inflatable penile prosthesis due to autoinflation.
A 59-year-old man presented with erosion of an inflatable penile prosthesis. The cylinders were not deflated and the right one protruded from the urethra. Autoinflation that causes failure of deflation is an annoying and embarrassing mechanical failure. This failure may cause intraurethral erosion of the prosthesis. Urologists should inform their patients who hope for implantation of an inflata...
متن کاملAutoinflation Leading to Failure of Two-Piece Ambicor Implantable Penile Prosthesis: An Outcome from a Methodical Treatment of Recalcitrant Stuttering Priapism
Introduction. We present the case of a patient who received a two-piece Ambicor penile prosthesis for idiopathic recurrent "stuttering" priapism refractory to other treatment options. The patient returned unable to deflate the device due to an interesting anatomically induced mechanical failure. Aims. To describe the method and findings of this inflatable prosthesis failure. Results. Prosthesis...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Canadian journal of plastic surgery = Journal canadien de chirurgie plastique
دوره 14 4 شماره
صفحات -
تاریخ انتشار 2006