Geometric Parameters Estimation and Calibration in Cone-Beam Micro-CT
نویسندگان
چکیده
The quality of Computed Tomography (CT) images crucially depends on the precise knowledge of the scanner geometry. Therefore, it is necessary to estimate and calibrate the misalignments before image acquisition. In this paper, a Two-Piece-Ball (TPB) phantom is used to estimate a set of parameters that describe the geometry of a cone-beam CT system. Only multiple projections of the TPB phantom at one position are required, which can avoid the rotation errors when acquiring multi-angle projections. Also, a corresponding algorithm is derived. The performance of the method is evaluated through simulation and experimental data. The results demonstrated that the proposed method is valid and easy to implement. Furthermore, the experimental results from the Micro-CT system demonstrate the ability to reduce artifacts and improve image quality through geometric parameter calibration.
منابع مشابه
Accurate technique for complete geometric calibration of cone-beam computed tomography systems.
Cone-beam computed tomography systems have been developed to provide in situ imaging for the purpose of guiding radiation therapy. Clinical systems have been constructed using this approach, a clinical linear accelerator (Elekta Synergy RP) and an iso-centric C-arm. Geometric calibration involves the estimation of a set of parameters that describes the geometry of such systems, and is essential...
متن کاملSelf-calibration of geometric and radiometric parameters for cone-beam computed tomography
Thanks to the advances in parallel processing hardware, iterative algorithms for cone beam reconstruction are now available with computation times acceptable for clinical use. At the same time they are able to accomodate more accurately the physical effects underlying the X-Ray imaging process. Many parameters are involved, which need to be precisely calibrated in order to achieve an accurate 3...
متن کاملA Calibration Method for Misaligned Scanner Geometry in Cone-beam Computed Tomography
Computed Tomography (CT) images often suffer from artifacts caused by misaligned scanner geometry of CT system. Calibration and correction must be done before image reconstruction. A method for calibration of misaligned scanner geometry in cone-beam CT with single-circle orbit is proposed. In this method, a four-point phantom is used to estimate a set of parameters that describe the geometry of...
متن کاملGeometric calibration of a stationary digital breast tomosynthesis system based on distributed carbon nanotube X-ray source arrays
Stationary digital breast tomosynthesis (sDBT) with distributed X-ray sources based on carbon nanotube (CNT) field emission cathodes has been recently proposed as an approach that can prevent motion blur produced by traditional DBT systems. In this paper, we simulate a geometric calibration method based on a proposed multi-source CNT X-ray sDBT system. This method is a projection matrix-based a...
متن کاملA geometrical calibration method for the PIXSCAN micro-CT scanner
Reconstruction in Cone-Beam Tomography can suffer from artifacts due to geometrical misalignments of the source-detector system. They can be avoided by a complete and precise description of the system. We present a high precision method for the geometric calibration for the PIXSCAN, a small animal X-ray CT scanner demonstrator based on hybrid pixel detectors (XPAD2). The specificities of the XP...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 15 شماره
صفحات -
تاریخ انتشار 2015