Note: calibration of atomic force microscope cantilevers using only their resonant frequency and quality factor.

نویسندگان

  • John E Sader
  • James R Friend
چکیده

A simplified method for calibrating atomic force microscope cantilevers was recently proposed by Sader et al. [Rev. Sci. Instrum. 83, 103705 (2012); Sec. III D] that relies solely on the resonant frequency and quality factor of the cantilever in fluid (typically air). This method eliminates the need to measure the hydrodynamic function of the cantilever, which can be time consuming given the wide range of cantilevers now available. Using laser Doppler vibrometry, we rigorously assess the accuracy of this method for a series of commercially available cantilevers and explore its performance under non-ideal conditions. This shows that the simplified method is highly accurate and can be easily implemented to perform fast, robust, and non-invasive spring constant calibration.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Note: Improved calibration of atomic force microscope cantilevers using multiple reference cantilevers.

Overall precision of the simplified calibration method in J. E. Sader et al., Rev. Sci. Instrum. 83, 103705 (2012), Sec. III D, is dominated by the spring constant of the reference cantilever. The question arises: How does one take measurements from multiple reference cantilevers, and combine these results, to improve uncertainty of the reference cantilever's spring constant and hence the overa...

متن کامل

Spring constant calibration of atomic force microscope cantilevers of arbitrary shape.

The spring constant of an atomic force microscope cantilever is often needed for quantitative measurements. The calibration method of Sader et al. [Rev. Sci. Instrum. 70, 3967 (1999)] for a rectangular cantilever requires measurement of the resonant frequency and quality factor in fluid (typically air), and knowledge of its plan view dimensions. This intrinsically uses the hydrodynamic function...

متن کامل

Variations in properties of atomic force microscope cantilevers fashioned from the same wafer.

Variations in the mechanical properties of nominally identical V-shaped atomic force microscope (AFM) cantilevers sourced from the same silicon nitride wafer have been quantified by measuring the spring constants, resonant frequencies and quality factors of 101 specimens as received from the manufacturer using the thermal spectrum method of Hutter and Bechhoefer. The addition of thin gold coati...

متن کامل

Normal and torsional spring constants of atomic force microscope cantilevers

Two methods commonly used to measure the normal spring constants of atomic force microscope cantilevers are the added mass method of Cleveland et al. @J. P. Cleveland et al., Rev. Sci. Instrum. 64, 403 ~1993!#, and the unloaded resonance technique of Sader et al. @J. E. Sader, J. W. M. Chon, and P. Mulvaney, Rev. Sci. Instrum. 70, 3967 ~1999!#. The added mass method involves measuring the chang...

متن کامل

Calibration of rectangular atomic force microscope cantilevers

A method to determine the spring constant of a rectangular atomic force microscope cantilever is proposed that relies solely on the measurement of the resonant frequency and quality factor of the cantilever in fluid ~typically air!, and knowledge of its plan view dimensions. This method gives very good accuracy and improves upon the previous formulation by Sader et al. @Rev. Sci. Instrum. 66, 3...

متن کامل