Upper limit of vulnerability during defibrillator implantations predicts the occurrence of appropriate shock therapy for ventricular fibrillation.
نویسندگان
چکیده
BACKGROUND The utility of the upper limit of vulnerability (ULV) test in patients undergoing defibrillator implantation has been reported, so the purpose of this study was to evaluate the difference in the clinical outcomes between patients with ULV ≤15 J or >15 J. METHODS AND RESULTS A total of 165 patients receiving an implantable cardioverter-defibrillator underwent a vulnerability test. At the time of the implantation, we delivered a 15-J shock on the T-peak and ±20 ms later to cover the most vulnerable part of the cardiac cycle. The clinical outcomes were prospectively analyzed. A 15-J shock induced ventricular fibrillation (VF) in 30 patients (ULV >15 J) and did not in 135 (ULV ≤15 J). The characteristics of the 2 groups were comparable. After a mean follow-up of 757 days, Kaplan-Meier curve analysis showed that the ULV ≤15 J group experienced less VF than the ULV >15 J group (log-rank P=0.003). The occurrence of ventricular tachycardia was similar between the 2 groups (P=0.140). Furthermore, the effectiveness of ATP was comparable. After adjusting for other known predictors of shock therapy, a ULV >15 J was independently associated with the occurrence of VF (hazard ratio: 6.25; 95% confidence interval: 1.913-20.40; P<0.01). CONCLUSIONS A high ULV value was associated with a high incidence of VF, which suggests that cardiac vulnerability to electrical shock may be linked to electrical instability.
منابع مشابه
Implantation of cardioverter defibrillators without induction of ventricular fibrillation.
BACKGROUND The upper limit of vulnerability (ULV) is the weakest shock at which ventricular fibrillation (VF) is not induced by a T-wave shock. This study tested the hypothesis that a vulnerability safety margin based on the ULV can be used as an implantable cardioverter-defibrillator implantation criterion. METHODS AND RESULTS Implantable cardioverter-defibrillators were implanted in 80 pati...
متن کاملInductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).
BACKGROUND Implantable cardioverter-defibrillators and cardiac resynchronization therapy defibrillators have relied on multiple ventricular fibrillation (VF) induction/defibrillation tests at implantation to ensure that the device can reliably sense, detect, and convert VF. The ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduc...
متن کاملUpper Limit of Vulnerabilit Reliably Predicts the Defibrillation Threshold in Humans
Background The upper limit of vulnerability is the stimulus strength above which electrical stimulation cannot induce ventricular fibrillation even when the stimulus occurs during the vulnerable period of the cardiac cycle. The purpose of this study was to test the hypothesis that the upper limit of vulnerability can accurately predict the defibrillation threshold in patients undergoing implant...
متن کاملRelation between upper limit of vulnerability and defibrillation threshold in humans.
BACKGROUND In the canine model, an upper limit of shock strength exists that can induce ventricular fibrillation during the vulnerable period of the cardiac cycle. This shock strength (the upper limit of vulnerability) closely correlates with the defibrillation threshold and supports the "upper limit of vulnerability" hypothesis of defibrillation. It is not known whether an upper limit of vulne...
متن کاملUpper limit of vulnerability reliably predicts the defibrillation threshold in humans.
BACKGROUND The upper limit of vulnerability is the stimulus strength above which electrical stimulation cannot induce ventricular fibrillation even when the stimulus occurs during the vulnerable period of the cardiac cycle. The purpose of this study was to test the hypothesis that the upper limit of vulnerability can accurately predict the defibrillation threshold in patients undergoing implant...