Comparison of a Novel, Single-Lead Atrial Sensing System With a Dual-Chamber Implantable Cardioverter- Defibrillator System in Patients Without Antibradycardia Pacing Indications

نویسنده

  • Christian Sticherling
چکیده

Background—Supraventricular tachyarrhythmias are the main cause for inappropriate therapy by implantable cardioverterdefibrillators (ICDs). For better rhythm discrimination, an atrial electrogram is helpful and usually obtained from an additional atrial lead, even in the absence of sinus node or atrioventricular nodal disease. An A -ICD system with integrated atrial sensing rings mounted 15 to 18 cm from the tip of an ICD lead may obviate the need to implant a separate atrial lead. The aim of the study was to compare the novel A -ICD and a conventional dual-chamber (DR)-ICD. Methods and Results—Two hundred forty-nine patients with standard ICD indications but no requirement for antibradycardia pacing were randomized to receive an A -ICD (n 124) or a DR-ICD (n 125). Implantation details, need for ICD system revision, long-term sensing, documented arrhythmia episodes, and the respective rhythm discrimination during follow-up were analyzed. The implantation time was significantly shorter in the A -ICD group (67 30 vs 79 30 minutes, P 0.003). Mean P-wave amplitudes were 3.5 0.8 mV (A -ICD) and 3.2 0.6 mV (DR-ICD) and remained stable during the follow-up period of 12 months. Surgical revision was necessary in 13 patients in the DR-ICD and 10 in the A -ICD group. All 593 ventricular tachyarrhythmia episodes were correctly discriminated. Sensitivity and specificity of supraventricular tachyarrhythmia discrimination were not different between the study groups. Conclusions—The novel A -ICD system can be implanted faster and is equivalent to a standard DR-ICD with regard to the detection of ventricular tachyarrhythmias and supraventricular tachyarrhythmias. It represents a useful alternative to obtain atrial sensing. Clinical Trial Registration—URL: http://www.clinicaltrials.gov. Unique identifier: NCT00324662. (Circ Arrhythm Electrophysiol. 2011;4:56-63.)

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of a novel, single-lead atrial sensing system with a dual-chamber implantable cardioverter-defibrillator system in patients without antibradycardia pacing indications: results of a randomized study.

BACKGROUND Supraventricular tachyarrhythmias are the main cause for inappropriate therapy by implantable cardioverter-defibrillators (ICDs). For better rhythm discrimination, an atrial electrogram is helpful and usually obtained from an additional atrial lead, even in the absence of sinus node or atrioventricular nodal disease. An A+-ICD system with integrated atrial sensing rings mounted 15 to...

متن کامل

"Two for the Price of One": A Single-Lead Implantable Cardioverter-Defibrillator System with a Floating Atrial Dipole.

In patients known to be a high risk for sudden cardiac arrest, implantable cardioverter defibrillators (ICD) are a proven therapy to reduce risk of death. However, in patients without conventional indications for pacing, the optimal strategy for type of device, dual- versus single-chamber, remains debatable. The benefit of prophylactic pacing in this category of patients has never been document...

متن کامل

Reduced risk for inappropriate implantable cardioverter-defibrillator shocks with dual-chamber therapy compared with single-chamber therapy: results of the randomized OPTION study.

OBJECTIVES The OPTION (Optimal Anti-Tachycardia Therapy in Implantable Cardioverter-Defibrillator Patients Without Pacing Indications) trial sought to compare long-term rates of inappropriate shocks, mortality, and morbidity between dual-chamber and single-chamber settings in implantable cardioverter-defibrillators (ICDs) patients. BACKGROUND The use of dual-chamber ICDs potentially allows be...

متن کامل

Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial.

CONTEXT Implantable cardioverter defibrillator (ICD) therapy with backup ventricular pacing increases survival in patients with life-threatening ventricular arrhythmias. Most currently implanted ICD devices provide dual-chamber pacing therapy. The most common comorbid cause for mortality in this population is congestive heart failure. OBJECTIVE To determine the efficacy of dual-chamber pacing...

متن کامل

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل