Performance evaluation of microPET: a high-resolution lutetium oxyorthosilicate PET scanner for animal imaging.
نویسندگان
چکیده
UNLABELLED A new dedicated PET scanner, microPET, was designed and developed at the University of California, Los Angeles, for imaging small laboratory animals. The goal was to provide a compact system with superior spatial resolution at a fraction of the cost of a clinical PET scanner. METHODS The system uses fiberoptic readout of individually cut lutetium oxyorthosilicate (LSO) crystals to achieve high spatial resolution. Each microPET detector consists of an 8 x 8 array of 2 x 2 x 10-mm LSO scintillation crystals that are coupled to a 64-channel photomultiplier tube by optical fibers. The tomograph consists of 30 detectors in a continuous ring with a 17.2-cm diameter and fields of view (FOVs) of 11.25 cm in the transaxial direction and 1.8 cm in the axial direction. The system has eight crystal rings and no interplane septa. It operates exclusively in the three-dimensional mode and has an electronically controlled bed that is capable of wobbling with a radius of 300 microm. We describe the performance of the tomograph in terms of its spatial, energy and timing resolution, as well as its sensitivity and counting-rate performance. We also illustrate its overall imaging performance with phantom and animal studies that demonstrate the potential applications of this device to biomedical research. RESULTS Images reconstructed with three-dimensional filtered backprojection show a spatial resolution of 1.8 mm at the center of the FOV (CFOV), which remains <2.5 mm for the central 5 cm of the transaxial FOV. The resulting volumetric resolution of the system is <8 microL. The absolute system sensitivity measured with a 0.74 MBq (20 microCi) 68Ge point source at the CFOV is 5.62 Hz/kBq. The maximum noise equivalent counting rate obtained with a 6.4-cm diameter cylinder spanning the central 56% of the FOV is 10 kcps, whereas the scatter fraction is 37% at the CFOV for an energy window of 250-650 keV and the same diameter cylinder. CONCLUSION This is the first PET scanner to use the new scintillator LSO and uses a novel detector design to achieve high volumetric spatial resolution. The combination of imaging characteristics of this prototype system (resolution, sensitivity, counting-rate performance and scatter fraction) opens up new possibilities in the study of animal models with PET.
منابع مشابه
Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.
UNLABELLED The microPET Focus is the latest generation microPET system dedicated to high-resolution animal imaging and incorporates several changes to enhance its performance. This study evaluated the basic performance of the scanner and compared it with the Primate (P4) and Rodent (R4) models. METHODS The system consists of 168 lutetium oxyorthosilicate (LSO) detectors arranged in 4 contiguo...
متن کاملEvaluation of Cerium Doped Lutetium Oxyorthosilicate ( LSO ) Scintillation Crystal for PET
A new scintillation crystal, cerium-doped lutetium oxyorthosilicate &SO) was recently discovered a, with light intensity 75 % of NaI(Tl), scintillation decay time of 12 ns (30%) and 42 ns (70%), effective Z of 66, and density of 7.4 glcc. The fast decay time and scintillation light output of LSO are superior to those of BGO for PET and the stopping power of LSO for 511 keV photons is only sligh...
متن کاملPerformance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.
UNLABELLED PET combined with CT has proven to be a valuable multimodality imaging device revealing both functional and anatomic information. Although PET/CT has become completely integrated into routine clinical application and also has been used in small-animal imaging, CT provides only limited soft-tissue contrast and, in preclinical studies, exposes the animal to a relatively high radiation ...
متن کاملA Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging.
UNLABELLED We developed a prototype small-animal PET scanner based on depth-encoding detectors using dual-ended readout of small scintillator elements to produce high and uniform spatial resolution suitable for imaging the mouse brain. METHODS The scanner consists of 16 tapered dual-ended-readout detectors arranged in a 61-mm-diameter ring. The axial field of view (FOV) is 7 mm, and the trans...
متن کاملPerformance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards.
UNLABELLED The aim of this work was to evaluate the performance characteristics of a newly developed dedicated breast PET scanner, according to National Electrical Manufacturers Association (NEMA) NU 4-2008 standards. METHODS The dedicated breast PET scanner consists of 4 layers of a 32 × 32 lutetium oxyorthosilicate-based crystal array, a light guide, and a 64-channel position-sensitive phot...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of nuclear medicine : official publication, Society of Nuclear Medicine
دوره 40 7 شماره
صفحات -
تاریخ انتشار 1999