ALMA Memo 461 The Calibration System revisited

نویسنده

  • A. Bacmann
چکیده

Different strategies of use of the polarization grids combined with two temperature loads for calibration are explored. It is shown that a system with an ambient load, a hot (100 • C or so) load, and a semi-transparent grid with 50 % transmission allows to reach a 1 % accuracy on the receiver gain even in presence of saturation. The best precision requires a device which offers the 5 major combinations: Sky, Sky+Ambient, Sky+Hot, Ambient, Hot. If saturation is negligible, the system can be used as a standard dual load calibration device, with better accuracy.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Dual - Load Calibration device revisited

Following ALMA memo 371, I re-investigate the dual load calibration scheme which has been proposed for ALMA. This memo improves on the requirements derived in memo 371 by using a better prediction for the atmospheric fluctuations at short time scales. It is shown that because of atmospheric properties and of the antenna geometry, a dual-load system located in the ALMA antenna subreflectors need...

متن کامل

EVLA Memo: #139 Plan for Imaging Algorithm Research and Development

Many scientific deliverables of the next generation radio telescopes require widefield imaging or high dynamic range imaging, or both. Some next generation telescopes nearing completion (EVLA and ALMA) indeed promise at least a ten-fold increase in the sensitivity compared to the existing telescopes. High imaging dynamic range requires removal of systematic errors to high accuracy and for long ...

متن کامل

Saturation by Noise and CW Signals in SIS Mixers

In ALMA Memo 321, Plambeck points out that saturation (gain compression) is likely to be a significant factor limiting the calibration accuracy of ALMA observations. In this memo, saturation by broadband noise and CW signals is analyzed for representative SIS receivers operating at different frequencies. Many SIS mixers in current use are expected to exhibit a significant degree of gain compres...

متن کامل

EVLA Memo 177 Polarimetric calibration and dynamic range issues

This memo considers some dynamic range issues resulting from limitations in the polarimetric calibration solutions used in typical observations. Whereas we focus on “traditional” radio interferometric telescopes such as the VLA, ATCA or ALMA, the analysis approach can be used for other interferometer arrays. The memo is mostly concerned with the principles: rather than addressing real data, thi...

متن کامل

ALMA Memo #303 Water Vapour Radiometers for ALMA

The JCMT-CSO single-baseline interferometer was the first to demonstrate phase correction using the 183 GHz line, using equipment built by Martina Wiedner, Richard Hills and colleagues. Only a limited quantity of data were gathered but the results (ALMA memo 252) were encouraging and suggested that even an uncooled system could provide effective phase calibration at submillimetre wavelengths. S...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2003