A hip wear simulator with 100 test stations.
نویسنده
چکیده
A novel high-capacity hip wear simulator of the pin-on-disc type was designed, built, and validated. This so-called Super-CTPOD (circularly translating pin-on-disc) device has as many as 100 separate test stations, being an advanced version of the previously validated 12-station CTPOD. A validity test was done so that in all stations the specimens and the test conditions were as similar as possible. Hence, for the first time in this field, an adequate number of similar tests was done for a proper statistical analysis of wear data. The pins were conventional, gamma-sterilized ultra-high molecular weight polyethylene, and the discs were polished CoCr. The lubricant was diluted calf serum and the test length 3 million cycles. In the course of the test, the pins became highly polished, whereas the discs remained practically unchanged. The majority of the polyethylene wear particles were rounded, with a mean diameter of 0.25 microm. The 100 wear factor values computed from the 100 steady state wear rate values of the pins were normally distributed, the mean +/- 95 per cent confidence interval being 1.63 +/- 0.017 x 10(-6) mm3 /N m. The standard deviation was 5.4 per cent of the mean. There were no outliers. The wear mechanisms and the wear factor agreed well with clinical findings. Altogether, the Super-CTPOD test system was shown to be a unique combination of validity, low variation, capacity, efficiency, reliability, productivity, economy, ease of operation, and compact size.
منابع مشابه
Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings
Hip joint simulators are used extensively for preclinical testing of hip replacements. The variation in simulator design and test conditions used worldwide can affect the tribological performance of polyethylene. The aim of this study was to assess the effects of simulator mechanics and design on the wear and creep of ultra-high-molecular-weight polyethylene. In the first part of this study, an...
متن کاملWear Testing of Total Hip Joints According to Iso and Development of a New Hip Joint Simulator According to Iso
INTRODUCTION The development of new, highly sophisticated materials for the articulating surfaces of total hip endoprostheses has led to an increasing demand for testing and verification. This includes tests and research regarding bio-compatibility, static load bearing capacity of the prostheses as well as tribological investigations. Since it has been found that wear debris play a dominant rol...
متن کاملAcetabular Cups in 60 mm Metal-on-Metal Bearings Subjected to Dynamic Edge-Loading with 70 Peak-Inclination in 10-Million Cycle Simulator Study
Wear simulation of total-hip arthroplasty (THA) involves hip biomechanics, tribology, bearing designs and cup wear-patterns. This is the first demonstration of cup edge-loading using the “Inverted-cup” test mode. Benefits included, (i) clinically relevant wear-patterns, and (ii) cup inclinations varying from ideal to edge-loaded during each 1-s simulator cycle. The 60 mm head and cup bearings i...
متن کاملWear of Crosslinked Uhmwpe in a Hip Joint Simulator
Crosslinking has been extensively introduced to reduce the wear of UHMWPE. In this study the wear rates and wear surfaces of UHMWPE with different levels of crosslinking were compared in a hip joint simulator. The UHMWPE was either non-irradiated, gamma irradiated in air (2.5MRad), or highly crosslinked (7.5 and 10MRad). The intentionally crosslinked materials showed a significantly lower volum...
متن کاملMicroseparation and stripe wear in alumina-on-alumina hip implants.
The combination of materials that still has highest wear resistance for total hip replacement is ceramic-on-ceramic. However, brittleness is a major concern for ceramics: in vivo and in vitro studies on ceramic hip prostheses correlate microseparation with hip noise, ceramic wear, or ceramic liner damage. Ceramic microseparation can lead to edge load, ceramic head wear, and squeaking. The aim o...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
دوره 219 5 شماره
صفحات -
تاریخ انتشار 2005