Free-air chamber correction factors for electron loss, photon scatter, fluorescence and bremsstrahlung
نویسنده
چکیده
Values for the correction factors for electron loss ke and photon scatter ksc were calculated by Burns [1] using the Monte Carlo code EGS4 [2], for all of the free-air chamber standards in use worldwide at that time. Spectral values were derived for the CCRI reference radiation qualities for lowand medium-energy x radiation [3]. In general, the calculated values were in close agreement with the values used at present by each NMI. The most notable discrepancy was in the values for ksc at the lower-energy qualities, where the calculated values were closer to unity by typically 0.2 %. This was attributed to the lack of inclusion of x-ray fluorescence transport in the standard EGS4 code.
منابع مشابه
Measurement Of Scatter Factors For Small Photon Fields Using Gaf chromic EBT2 Film
Introduction: Small field dosimetry is challenging for radiotherapy dosimetry. measurement of the output factor in the air and water (Sc, Scp) is one of the input parameters for commissioning of treatment planning systems and beam modeling. The aims of this study are to measured Sc,Scp for small fields with EBT2 and Ion chamber and design a appropriate mini-phantom for small fi...
متن کاملAn assessment of the Photon Contamination due to Bremsstrahlung Radiation in the Electron Beams of a NEPTUN 10PC Linac using a Monte Carlo Method
Introduction: In clinical electron beams, most of bremsstrahlung radiation is produced by various linac head structures. This bremsstrahlung radiation dose is influenced by the geometry and construction of every component of the linac treatment head structures. Thus, it can be expected that the amount of the contaminated photon dose due to bremsstrahlung radiation varies among different linacs,...
متن کاملEstimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods
Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...
متن کاملDetermination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation
Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method usin...
متن کاملEstimation of bremsstrahlung photon fluence from aluminum by artificial neural network
Background: As bremsstrahlung photon beam fluence is important parameter to be known in a photonuclear reaction experiment as the number of produced particle is strongly depends on photon fluence. Materials and Methods: Photon production yield from different thickness of aluminum target has been estimated using artificial neural network (ANN) model. Target thickness and incoming electr...