Advantages of the Beall valve prosthesis.
نویسندگان
چکیده
Followup of 29 patients with a Beau mitral valve prosthesii has shown important advantages over the KayShiiey valve. No late deaths and no emboli were noted during an average 14 month (range 8 to 27 months) follewup. Mitral insufficiency was not noted clinically or by angiography. Hernodynamic studies in nine patients 16 months postoperation showed improvement. Although all patients had evidence of prosthetic obstruction postoperatively, particularly during exercise, good symp tomatic improvement, lack of emboli and preservation of disc motion, reflect advantages of thii valve over others.
منابع مشابه
Ten-year pathological study of Beall mitral valve prostheses.
Detailed pathological studies are reported on 20 cloth-covered Beall mitral valve prostheses, which had to be removed for valve failure nine months to 10 years after implantation. Results in all patients except one indicated that there was a progressive and persistent granulomatous reaction that continued up to at least 10 years. Although the rate of tissue ingrowth into the valve decreased wit...
متن کاملThe pathology of wear in the Beall model 104 heart valve prosthesis.
We examined 13 Beall model 104 prostheses recovered at surgery or autopsy 10 to 84 1/2 months after insertion and observed the pattern of prosthesis wear. We defined wear as "mild" when the disc was notched but neither the metal of the struts was exposed not the cloth seat torn. In "moderate" wear the disc was notched, the Teflon coating of the struts was worn away exposing the underlying meta...
متن کاملMitral disc-valve variance.
This report deals with a rare complication of disc-valve prosthesis in the mitral area. A significant disc poppet and struts destruction of mitral BeaU valve prostheses occurred 20 and 17 months after implantation. The resulting valve incompetence in the first case contributed to the death of the patient. The durability of Teflon prosthetic valves appears to be in question and this type of valv...
متن کاملAN INNOVATIVE SURGIC AL TECHNIQUE FOR CONTINUOUS SUTURING OF PROSTHESIS IN MITR AL VALVE REPL A CEMENT
Mitral valve replacement (MVR) is usually performed in two methods of running and interrupted suturing. In running method, the suturing of the valve is usually with some traction on the annulus and also there is a risk of aortic valve injury. This study was conducted to evaluate a new technique for suturing to avoid these complications and decrease the time of cardiac arrest. One hundred an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Chest
دوره 59 3 شماره
صفحات -
تاریخ انتشار 1971