Persistence in NICMOS: Results from Thermal Vacuum Data
نویسندگان
چکیده
This ISR presents an analysis of the System Level Thermal Vacuum (SLTV) persistence tests performed with NICMOS. In August 1996, persistence data was obtained as part of the SLTV, and further data were obtained in December 1996 after modifications had been made to the instrument operating procedures. The goal was to determine if NICMOS exhibits an “after-glow” or “persistence” when read out in the dark immediately after having been exposed to bright illumination; and then to characterize the behavior of such persistence. The results and analysis of the SLTV tests show that the NICMOS detectors do indeed exhibit a significant level of persistence. The level of persistence seems to decay following a power law. Furthermore, the comparison of data observed with different AUTO-FLUSH patterns verifies the efficiency of the procedure when AUTO-FLUSH is turned “on”. Quantitative analysis of the on-orbit performance of the NICMOS detectors with regard to persistence will be described in a later ISR.
منابع مشابه
The Polarimetric Capabilities of NICMOS
The polarimetric capabilities of NICMOS are demonstrated from data obtained during the Early Release Observations of IRC +10216 and CRL 2688 (the Egg Nebula). Preflight Thermal Vacuum tests revealed that each polarizer has a unique polarizing efficiency, and that the position angle offsets differ from the nominal positions of 0◦, 120◦ & 240◦. Therefore an algorithm different from that of an ide...
متن کاملBright Earth Persistence in NICMOS
We report the presence of image persistence in NICMOS data due to prior saturation of the array by the bright Earth as well as an algorithm to remove its impact. BEP (Bright Earth Persistence) may occur when ACS and NICMOS are used in parallel, and the data dump of ACS delays the insertion of the NICMOS filter blank as long as 6 minutes beyond the Bright Earth avoidance angle. We have construct...
متن کاملNICMOS Pointed Thermal Background: Results from On-Orbit data
This ISR presents the results of the analysis of NICMOS pointed thermal background data taken as part of the Servicing Mission Orbital Verification (SMOV). Observations in NICMOS Camera 3 are taken at different pointings and analysis of the data shows the HST thermal emission to be fairly stable. Furthermore, data obtained with multiple filters shows no significant differences with the backgrou...
متن کاملAnalysis of Polarized Light with NICMOS
The Near Infrared Camera and Multi-Object Spectrometer (NICMOS) aboard the Hubble Space Telescope (HST ) contains optics in Cameras 1 and 2 (NIC1 and NIC2) that enable imaging polarimetry at D1 and 2 km, respectively, with unprecedented spatial detail. PreÑight thermal vacuum tests revealed that the three polarizers in each camera have unique polarizing efficiencies and that the rotational orie...
متن کاملRemoving Post-SAA Persistence in NICMOS Data
A general concern with NICMOS images is the issue of “persistence”, or residual charge from bright sources and cosmic rays that remains trapped in pixels and is gradually released during subsequent exposures. This is particularly noticeable after HST transits the South Atlantic Anomaly (SAA), where the cosmic ray rate is so high that their residual flux contributes a significant noise component...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 1999