Impact of carotid stent cell design on vessel scaffolding: a case study comparing experimental investigation and numerical simulations.

نویسندگان

  • Michele Conti
  • Denis Van Loo
  • Ferdinando Auricchio
  • Matthieu De Beule
  • Gianluca De Santis
  • Benedict Verhegghe
  • Stefano Pirrelli
  • Attilio Odero
چکیده

PURPOSE To quantitatively evaluate the impact of carotid stent cell design on vessel scaffolding by using patient-specific finite element analysis of carotid artery stenting (CAS). METHODS The study was organized in 2 parts: (1) validation of a patient-specific finite element analysis of CAS and (2) evaluation of vessel scaffolding. Micro-computed tomography (CT) images of an open-cell stent deployed in a patient-specific silicone mock artery were compared with the corresponding finite element analysis results. This simulation was repeated for the closed-cell counterpart. In the second part, the stent strut distribution, as reflected by the inter-strut angles, was evaluated for both cell types in different vessel cross sections as a measure of scaffolding. RESULTS The results of the patient-specific finite element analysis of CAS matched well with experimental stent deployment both qualitatively and quantitatively, demonstrating the reliability of the numerical approach. The measured inter-strut angles suggested that the closed-cell design provided superior vessel scaffolding compared to the open-cell counterpart. However, the full strut interconnection of the closed-cell design reduced the stent's ability to accommodate to the irregular eccentric profile of the vessel cross section, leading to a gap between the stent surface and the vessel wall. CONCLUSION Even though this study was limited to a single stent design and one vascular anatomy, the study confirmed the capability of dedicated computer simulations to predict differences in scaffolding by open- and closed-cell carotid artery stents. These simulations have the potential to be used in the design of novel carotid stents or for procedure planning.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Patient-specific Finite Element Analysis of Carotid Artery Stenting: Impact of constitutive vessel modeling on vessel wall stress distribution

Carotid artery stenting (CAS) is emerged as a safe and cost-effective treatment of carotid artery (CA) stenosis [1]. However the long-term efficacy of CAS is under clinical evaluation and, in particular, the role of in-stent restenosis (ISR) is not clear. More data and dedicated studies are necessary to elucidate the mechanisms of ISR and their relation with the novel carotid stent designs. Exp...

متن کامل

Carotid artery stenting simulation: from patient-specific images to finite element analysis.

The outcome of carotid artery stenting (CAS) depends on a proper selection of patients and devices, requiring dedicated tools able to relate the device features with the target vessel. In the present study, we use finite element analysis to evaluate the performance of three self-expanding stent designs (laser-cut open-cell, laser-cut closed-cell, braided closed-cell) in a carotid artery (CA). W...

متن کامل

Virtual scaffolding and radial strength testing of stents

Virtual design tools are gaining more and more importance in the development and evaluation of medical devices. Inherently each medical device is subjected to (very) different (patient) specific loading conditions and thus requires dedicated simulation strategies. In addition, specific post-processing of the data obtained from the numerous simulations is needed to evaluate the device behavior. ...

متن کامل

Numerical investigation of stopping maneuver for two by reversing propeller at different speeds

Maneuverability is an important aspect of marine vehicle design. With the development and utilization of ocean resources, maritime transportation is becoming increasingly busy. Stopping ability has great effect on the safety of ship maneuvering for those large ships. Therefore, it is necessary to study the behavior of the ship during the stopping maneuver to ensure the safety of navigation. In ...

متن کامل

Experimental and numerical crashworthiness investigation of hybrid composite aluminum tubes under dynamic axial and oblique loadings

This research deals with axial and oblique impact crash tests on simple and hybrid composite tubes. Axial and oblique impact tests have been generated on simple and hybrid composite tubes with one, two and three layers. A drop test rig was used to conduct the experiments. Furthermore, in order to gain more detailed knowledge about the crash process, finite element simulations of the experim...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists

دوره 18 3  شماره 

صفحات  -

تاریخ انتشار 2011