Standard Reference Material (SRM 1990) For Single Crystal Diffractometer Alignment

نویسندگان

  • W. Wong-Ng
  • T. Siegrist
  • G. T. DeTitta
  • L. W. Finger
  • H. T. Evans
  • E. J. Gabe
  • G. D. Enright
  • J. T. Armstrong
  • M. Levenson
  • L. P. Cook
  • C. R. Hubbard
چکیده

An international project was successfully completed which involved two major undertakings: (1) a round-robin to demonstrate the viability of the selected standard and (2) the certification of the lattice parameters of the SRM 1990, a Standard Reference Material(®) for single crystal diffractometer alignment. This SRM is a set of ≈3500 units of Cr-doped Al2O3, or ruby spheres [(0.420.011 mole fraction % Cr (expanded uncertainty)]. The round-robin consisted of determination of lattice parameters of a pair of crystals: the ruby sphere as a standard, and a zeolite reference to serve as an unknown. Fifty pairs of crystals were dispatched from Hauptman-Woodward Medical Research Institute to volunteers in x-ray laboratories world-wide. A total of 45 sets of data was received from 32 laboratories. The mean unit cell parameters of the ruby spheres was found to be a=4.7608 ű0.0062 Å, and c=12.9979 ű0.020 Å (95 % intervals of the laboratory means). The source of errors of outlier data was identified. The SRM project involved the certification of lattice parameters using four well-aligned single crystal diffractometers at (Bell Laboratories) Lucent Technologies and at NRC of Canada (39 ruby spheres), the quantification of the Cr content using a combined microprobe and SEM/EDS technique, and the evaluation of the mosaicity of the ruby spheres using a double-crystal spectrometry method. A confirmation of the lattice parameters was also conducted using a Guinier-Hägg camera. Systematic corrections of thermal expansion and refraction corrections were applied. These rubies- are rhombohedral, with space group [Formula: see text]. The certified mean unit cell parameters are a=4.76080±0.00029 Å, and c=12.99568 ű0.00087 Å (expanded uncertainty). These certified lattice parameters fall well within the results of those obtained from the international round-robin study. The Guinier-Hägg transmission measurements on five samples of powdered rubies (a=4.7610 ű0.0013 Å, and c = 12.9954 ű0.0034 Å) agreed well with the values obtained from the single crystal spheres.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Screening trace elements and electrolytes in serum by inductively-coupled plasma emission spectrometry.

This rapid, accurate procedure for trace elements and electrolytes in serum requires little sample preparation: to 1 mL of serum a single reagent is added that contains trichloroacetic and hydrochloric acids for protein precipitation, hydroxylamine sulfate for iron reduction, and yttrium as the internal standard. After centrifugation, the supernates are directly analyzed for Na, K, Mg, Ca, Pi, ...

متن کامل

Standard Reference Material 1751: Gallium Melting-Point Standard

The melting point of gallium (29.7646 °C) is a defining thermometric fixed point of the International Temperature Scale of 1990 (ITS-90). Realization of this melting point is performed using a fixedpoint cell containing high-purity (≥99.999 99 wt. % pure) gallium. A 30 kg single lot of gallium (99.999 9956 wt. % pure) constituting Standard Reference Material (SRM) 1751 was evaluated, and certif...

متن کامل

Neutron powder diffraction in materials with incoherent scattering: an illustration of Rietveld refinement quality from nondeuterated gypsum

The power of the state-of-the-art neutron powder diffractometer suite at the Institut Laue–Langevin for investigating the structure of nondeuterated materials is presented using gypsum, CaSO4 2H2O, as a reference material. It is shown that flexible modern neutron powder diffraction instruments at reactorbased sources can yield data with sufficient counting statistics above the incoherent scatte...

متن کامل

Frozen human serum reference material for standardization of sodium and potassium measurements in serum or plasma by ion-selective electrode analyzers.

Three interlaboratory round-robin studies (RR1, RR2, and RR3) were conducted to identify a serum-based reference material that would aid in the standardization of direct ion-selective electrode (ISE) measurements of sodium and potassium. Ultrafiltered frozen serum reference materials requiring no reconstitution reduced between-laboratory variability (the largest source of imprecision) more than...

متن کامل

Molar absorptivities of bilirubin (NIST SRM 916a) and its neutral and alkaline azopigments.

Three laboratories in the U.S. and two in the Netherlands determined molar absorptivities (epsilon) of Standard Reference Material (SRM) 916a Bilirubin from the National Institute of Standards and Technology. In caffeine reagent the average epsilon values were 50,060 and 48, 980 L.mol-1.cm-1 at 432 and 457 nm, respectively. The epsilon value of the blue azopigment, obtained with the Reference M...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 106  شماره 

صفحات  -

تاریخ انتشار 2001